NAVY ELECTRICITY AND ELECTRONICS TRAINING SERIES MODULE 16- TEST EQUIPMENT

NAVEDTRA 14188A · CHAPTER 1, 3, 6

Chapter 1 Test Equipment Administration and Use

p. 11

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-1 UNCLASSIFIED 1 TEST EQUIPMENT ADMINISTRATION AND USE LEARNING OBJECTIVES

Learning objectives are stated at the beginning of each chapter. These learning objectives serve as a preview of the information you are expected to learn in the chapter. The comprehensive check questions are based on the objectives. By successfully completing the NRTC, you indicate that you have met the objectives and have learned the information. The learning objectives are listed below.

1. Upon completing this chapter, you should be able to: 2. Describe the Ship Configuration and Logistic Support Information System (SCLSIS). 3. State the differences between calibration and repair. 4. Explain the various calibration status labels used by the Navy. 5. List the procedures for obtaining repairs to test equipment. 6. Describe the Metrology Automated System for Uniform Recall and Reporting (MEASURE) System and the purpose of the Metrology Equipment Recall and Reporting (METER) card and recall schedule. 7. Describe major test equipment references available to you. 8. Explain the purposes and benefits of testing. 9. State the safety precautions involved in working with test equipment. 10. List three precautions you should observe to avoid damaging electric measuring instruments. 11. State the correct procedures for using a safety shorting probe. 12. Describe resistance, voltage, and current measurements in terms of purposes, methods, and instruments used. 13. Describe how capacitance and inductance are measured. 14. Explain the operation of bridges in the measurement of unknown resistances, capacitances, and inductances.

p. 12

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-2 UNCLASSIFIED 1.1 INTRODUCTION One purpose of this chapter is to acquaint you with the practical use of test equipment. The presence of adequate test equipment in your shop is not in itself a "cure-all" for making repairs to complex electronic equipment. You must know how to best use the equipment available. First, however, you must understand the basis of electronic theory and be able to apply it to the system under repair.

Another purpose of this chapter is to introduce you to calibration and repair procedures, and basic voltage and current measurements. You will also learn how ac bridges are used for precise measurements of resistance, capacitance, and inductance.

Much of the theory of operation and practical applications of the basic types of test instruments used in electrical and electronic circuits are found in the instruction books and technical manuals that accompany various equipments. You should read and understand these books before you attempt to use any test instrument. You should also know the established safety precautions to ensure your safety and safe equipment operating procedures to protect equipment from damage.

1.2 TEST EQUIPMENT IDENTIFICATION One of the first things you must learn as a maintenance technician is how to identify the various electronic equipment and components by their appropriate nomenclatures. You will find that several methods are used to identify test equipment used; this may be somewhat confusing to you at first. For example, a Tektronix Model 541A oscilloscope can also be identified as a CBTV-541A. The Joint Electronics Type Designation System (JETDS) is used by all branches of the military to identify equipment by a system of standardized nomenclatures.

Q-1. What system is currently used by all branches of the military to identify test equipment?

p. 13

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-3 UNCLASSIFIED 1.3 ELECTRONIC TEST EQUIPMENT CLASSIFICATION The Electronic Test Equipment Classification Board was established in 1973 to control the increased use of undesirable electronic test equipment (ETE) in fleet and shore activities. The board classifies electronic test equipments as GENERAL PURPOSE (GPETE) or SPECIAL PURPOSE (SPETE) and assigns responsibility for their management. Items classified as general purpose are managed by the Space and Warfare Systems Command (SPAWARSYSCOM). Items classified as special purpose are managed by the individual systems command that generates the requirement.

GPETE is test equipment that has the capability, without modification, to generate, modify, or measure a range of parameters of electronic functions required to test two or more equipments or systems of basically different design.

Special-purpose electronic test equipment (SPETE) is specifically designed to generate, modify, or measure a range of parameters of electronic functions of a specific or peculiar nature required to test a single system or equipment. These special test equipments are procured by the systems command that has the responsibility for the system/equipment requiring the SPETE for maintenance.

Q-2. Name the two classes of test equipment. Q-3. What test equipment is designed to generate, modify, or measure a range of parameters of electronic functions of a specific nature required to test a single system or equipment? Until the ETE classification board was established, the uncontrolled increase in use of nonstandard GPETE had resulted in loss of inventory control and increased support costs. NESEA has the responsibility for evaluating requests to purchase nonstandard GPETE and for recommending its approval or disapproval to NAVSEA. NAVSEA will then forward its final decision to the originating command for such requests.

p. 14

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-4 UNCLASSIFIED 1.3.1 Ship Configuration and Logistic Information System (SCLSIS) Program The Navy must maintain, update, and calibrate thousands of pieces of equipment. To do this, the SHIP CONFIGURATION AND LOGISTIC SUPPORT INFORMATION SYSTEM (SCLSIS) program was designed to keep track of all installed and portable equipment in the fleet. SCLSIS is used to keep up with the existence, location, and changes made to equipment. The SCLSIS program seeks to improve the quality of equipment reporting, provide information needed by other Navy management systems, and reduce record keeping. It is also designed to assist Navy supply systems that furnish spares, documentation, and training necessary to support installed and portable equipment.

Therefore, the inventory of assigned test equipment on board ship is directly related to SCLSIS records. Properly maintained SCLSIS records also show the complete inventory of test equipment on board by quantity, serial number, and location. The SCLSIS program has two basic elements: (1) VALIDATION, to establish a baseline data inventory, and (2) INVENTORY UPDATING, to correct errors or omissions and to document configuration changes.

Q-4. Name the two basic elements of the SCLSIS program.

p. 15

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-5 UNCLASSIFIED 1.4 CALIBRATION AND REPAIR PROCEDURES The difference between the terms calibration and repair needs to be addressed before we proceed further. Calibration is little more than checking, adjusting, or systematically aligning a test instrument to a known standard. To do this, you must ensure that the equipment you send to the calibration lab is in working order.

The calibration lab is where actual repair work becomes important. Obvious problems, such as open power cords, burned components, broken meters, and missing hardware, should be repaired or replaced before sending equipment to the calibration lab. Most calibration labs with which you will deal will be part of an intermediate maintenance activity (IMA) on board a tender.

1.4.1 Calibration Status You can determine the calibration status of any test equipment by checking the calibration label or tag located on the equipment. These calibration labels or tags advise you as to whether the item is usable and within its calibration interval. Tags and labels to be used in the METROLOGY CALIBRATION (METCAL) coordination program are listed in the following paragraphs. No other calibration labels or tags are authorized to be placed on test equipment.

1.4.1.1 Calibrated Label The CALIBRATED label, shown in view A of figure 1-1, has black lettering and a white background and comes in two sizes. It is the most commonly used label in the METCAL program. This label indicates that the instrument to which it is attached is within its applicable tolerance on all parameters. If there are any qualifying conditions for use of the instrument, one of the other labels described in the next paragraphs should be used.

1.4.1.2 Calibrated - Refer to Report Label The CALIBRATED - REFER TO REPORT label, shown in view B of figure 1-1, has red lettering and a white back ground. It comes in two sizes and is used when you must know the actual measurement values to use the instrument.

Q-5. What calibration label is used when actual measurement values must be known to use the test equipment?

p. 16

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-6 UNCLASSIFIED

Figure 1-1 Calibration labels and tags

p. 17

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-7 UNCLASSIFIED 1.4.1.3 Special Calibration Labels Two SPECIAL CALIBRATION labels are shown in view C of figure 1-1 that have black lettering and a yellow background; the size and content of the labels are different. A SPECIAL CALIBRATION tag (figure 1-1, view C) is used with the smaller of the two labels. These labels or tag are used when some unusual or special condition in the calibration should be drawn to your attention.

Such special conditions may be deviations from usual calibration tolerances, multiple calibration intervals, or a requirement for in-place calibration. The special condition that resulted in the SPECIAL CALIBRATION label should be described on the large label when sufficient space is available on the instrument or on the tag when the small label is used. Brief descriptions of special conditions are provided in the following paragraphs.

Q-6. An instrument that must be calibrated in place requires what type of calibration label? In cases where you do not require full instrument capability, the calibration can be performed with reduced tolerances or cover less than all ranges and parameters. This approach is often used when the instrument does not meet full calibration tolerances on certain ranges or parameters, but can still meet user requirements. On the other hand, the special calibration may be for higher accuracy than usual on a short-term basis upon your specific request.

MULTIPLE CALIBRATION INTERVALS - Some instruments have components that require calibration less frequently than the rest of the instrument. For example, the attenuator in a signal generator may require calibration every 12 months, whereas the rest of the instrument parameters should be calibrated every 4 months. Since the attenuator calibration is time consuming and may require unavailable standards, use of the multiple- interval approach can save considerable time (man-hours) as well as permit the more frequent calibration to be performed at a lower level laboratory.

When a specific instrument has been designed for multiple calibration intervals, such information is provided in the applicable calibration procedure. The SPECIAL CALIBRATION label or tag is annotated with the words MULTIPLE INTERVAL, and the type of calibration performed is indicated; for example, partial 1 of 2, 2 of 2, complete calibration, and so forth. The calibration due date reflects the due date of the next partial or complete calibration.

CALIBRATION IN-PLACE - Some instruments should be calibrated in-place. Annotation on the SPECIAL CALIBRATION label or tag will alert both you and the calibrator that the instrument should not be removed, but should be calibrated in-place.

p. 18

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-8 UNCLASSIFIED 1.4.1.4 User Calibration Label Some test and measuring equipment (T&ME) should be calibrated by you instead of your referring the instrument to a calibration facility. For example, some instruments, such as hardness testers and densitometers, are provided with their own standards and should be calibrated each time used, or at least very frequently. Some instruments, such as oscillographic recorders, may require calibration before, during, and after each use.

Other automatic test equipment (ATE) has self-calibration tests that should be performed each time used or each day of use. Still other instruments are calibrated as part of checkout procedures performed daily or weekly and recorded in maintenance logs. Whenever recognized, the requirement for calibration by the user and the calibration interval (each use - daily, weekly, every 100 hours - each overhaul, and so forth) is indicated in the Metrology Requirements List (METRL).

The USER CALIBRATION label, shown in view D of figure 1-1, has black lettering and a white background and is affixed when the calibration is performed by the user; however, this label is not replaced at each calibration. When the label is first attached to the instrument, it is annotated as to the appropriate calibration interval. Records of calibrations performed, when other than each time used, should be by normal maintenance practices; that is, in the maintenance log, on maintenance action forms, and so forth.

1.4.1.5 Inactive - Calibrate Before Use Label In the event that an individual instrument due for recalibration will not be used for some time in the future, you may indefinitely postpone the recalibration by affixing an inactive label to the instrument. As shown in view E of figure 1-1, the INACTIVE - CALIBRATE BEFORE USE label has green lettering and a white background. The INACTIVE label remains on the instrument until it is recalibrated. The instrument is not to be used while bearing this label.

p. 19

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-9 UNCLASSIFIED 1.4.1.6 Calibration Not Required Label Test equipment standards and T&ME not requiring calibration are shown as CALIBRATION NOT REQUIRED. This label, shown in view F of figure 1-1, has orange letters and a white background. It is attached to and should remain on the instrument indefinitely unless its calibration requirements change. If the instrument is not listed in METRL, you should use the following criteria when placing instruments in the CALIBRATION NOT REQUIRED category:

• Instrument does not make quantitative measurements nor provide quantified outputs. • The device is "fail-safe"; that is, operation beyond specified tolerances will be apparent to the user. • All measurement/stimulus circuits are monitored during use by calibrated instruments or are dependent on external known or calibrated sources for performance within required limits. (When determining that an instrument falls into the CALIBRATION NOT REQUIRED category, you should annotate the label as to the authority for the decision, such as METRL, technical manual, letter or message from higher authority.)

1.4.1.7 Rejected - refer To Attached Tag Label In the event that an instrument fails to meet the acceptance criteria during calibration and cannot be adequately repaired, a REJECTED - REFER TO ATTACHED TAG label is placed on the instrument and all other servicing labels removed. This label, as shown in view G of figure 1-1, has black letters and a red background. In addition to the REJECTED label, a REJECTED tag, giving the reason for rejection and other information as required, is attached to the instrument. Both the label and tag remain on the instrument until it is repaired and recalibrated. The instrument is not to be used while bearing a REJECTED label.

1.4.1.8 Calibration Void If Seal Broken Label The CALIBRATION VOID IF SEAL BROKEN label, shown in view H of figure 1-1, has black letters and a white background. It is placed over readily accessible (usually exterior) adjustments to prevent tampering by the user when such tampering could affect the calibration. The label should not be used to cover adjustments or controls that are part of the normal use and operation of the instrument. This label may also be used to prevent removal and/or interchange of plug-ins, modules, subassemblies, and so forth, when such removal or interchange would affect the calibration.

p. 20

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-10 UNCLASSIFIED 1.4.2 Repair Procedures If you are unable to replace a known failed component with onboard spares, you can often locate the replacement component from other supply sources. The replacement component can then be delivered, along with the inoperative equipment, to the IMA. So by sending the repair part along with the equipment, you can reduce repair time considerably. This is particularly true when your unit is getting under way and no time is available for you to complete the repair before calibration. Most operational commands have a higher supply priority for purchase of repair parts than the IMA can use.

1.4.2.1 "No Reject" Policy IMAs have a "no reject" policy on test equipment to provide operational test equipment in a timelier manner. The "no reject" policy says, in effect, that test equipment submitted to the IMA for calibration, which is later found to require repair, will be repaired by the repair department of the IMA. Before this policy, any equipment found inoperative by the calibration lab was marked REJECTED, the reasons stated, and the equipment returned uncalibrated to the ship for repairs. The "no reject" policy does not relieve you of your responsibility to ensure your equipment is in working order prior to submitting it for calibration. Its purpose is to streamline the procedure and cut down delays in returning your equipment to you calibrated and ready to use.

1.4.2.2 Responsibility for Repair and Maintenance of Test Equipment Generally, the responsibility for repair and maintenance of test equipment is placed on maintenance personnel. In some cases, however, maintenance personnel are not authorized to make repairs. Then the test instrument must be sent to a shore repair/calibration facility.

Q-7. Responsibility for repair and maintenance of test equipment generally rests with what group of personnel? When test equipment is sent for calibration and repair, all accessories, such as probes, adapters, and calibration sheets, should be included. Only in emergencies or special situations should partial repair or calibration be attempted on test equipment designated as nonrepairable. Such emergency repairs should be noted on a tag attached to the unit and an entry made on the MEASURE card (discussed shortly). The equipment should then be sent at the earliest opportunity to an authorized facility so that permanent repairs can be made and the unit calibrated.

p. 21

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-11 UNCLASSIFIED 1.4.3 Stowage and Handling of Test Equipment Most electronic test equipment is precision equipment. Such equipment must be handled with care to properly perform its designed functions. Rough handling, excessive heat, moisture, and dust all affect the useful life of the equipment. Bumping or dropping a test instrument may ruin the calibration of a meter, cause short circuits, or damage electronic elements inside the case. Sharp bends, creases, or dents in coaxial test cables can alter the expected attenuating effect and cause false meter readings or measurements. Forced air cooling, dust filters, and heaters are used in many pieces of equipment. This test equipment requires clean air filters for proper ventilation and a warm-up period that permits units in the equipment to maintain calibrated standards.

Electronic test equipment should be stowed in a dry location with the dust cover (if provided) in place. Dust covers for spare plug-in units should be constructed for such stowage. For ease in performing maintenance, the test equipment should be stowed at a location convenient to equipment spaces. If possible, related test equipment should be mounted in the equipment spaces. This reduces the problem of finding adequate stowage space elsewhere.

In stowage spaces, individual pieces of test equipment should be held in place by stretch seat-belt-type straps. If bars are used to hold equipment on shelves, meters and control knobs should be protected by blocking the equipment to prevent it from rolling and sliding on the shelf. Test equipment too large for shelf stowage should be kept in stowage cases and tie-downs provided to secure the cases. Refer to Stowage Guide for Portable Test Equipment, NAVSEA ST000-AB-GYD-010/GPETE, to determine adequate stowage space and proper weight support requirements.

1.4.4 The Metrology Automated System for Uniform Recall and Reporting (Measure) For the sake of simplicity, we will use the more commonly used acronym MEASURE instead of the full name to describe this system in the next discussion.

MEASURE is a data processing system designed to provide a standardized system for the recall and scheduling of test, measurement, and diagnostic equipment (TMDE) into calibration facilities. It also provides for the documentation of data pertaining to the calibration actions performed by these facilities.

The primary reference document that describes the operation of the MEASURE system is Metrology Automated System for Uniform Recall and Reporting (MEASURE) Users Manual, OP 43P6A. The Chief of Naval Operations oversees this program and establishes policy and guidelines.

p. 22

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-12 UNCLASSIFIED Q-8. What Navy office oversees the MEASURE program? Each naval activity must ensure that the test equipment for which it has been assigned primary responsibility is submitted on a timely basis to a calibration activity for required calibration.

The MEASURE program is designed to assist these naval activities in the fulfillment of this responsibility. MEASURE does this by providing for the automatic scheduling and recall of all such test equipment for calibration.

Each activity submits an initial inventory, using the form shown in figure 1-2, to its Metrology Calibration Representative (METCALREP) for approval. The METCALREP then forwards the inventory to the Measure Operational Control Center (MOCC). The MOCC, based on the information contained on these inventory report forms, provides the necessary preprinted Metrology Equipment Recall and Reporting (METER) card. Figure 1-3 illustrates a MEASURE METER card.

Figure 1-2 MEASURE TMDF Inventory report form

p. 23

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-13 UNCLASSIFIED

Figure 1-3 MEASURE METER card

p. 24

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-14 UNCLASSIFIED In part, the METER card is preprinted with information taken from the initial inventory data submitted on the inventory report forms together with such updated data as may appear on any prior METER card. The remaining information required is entered on the card by the user of the equipment or the calibration activity, as appropriate.

The METER card is used to report changes, additions, or deletions to the user activity’s inventory. It is also used to report changes in custody of the item of test equipment. The procedure for filling out the METER card is outlined in the appendixes of the MEASURE Users Manual. Blank METER cards can be obtained through the responsible METCALREP.

A computer printout recall schedule is also generated by the MEASURE system. The purpose of this printout is to list those items of equipment that are due for calibration. Each recall schedule is composed of a set of four identical copies. One set is provided to the calibration activity as an aid to workload planning; a second set is sent to the user’s activity. The recall schedule is one of several products/formats sent automatically by the MEASURE Operation Control Center to the user activity on a regular basis. The MOCC automatically distributes the following products to user activities at the intervals shown:

DOCUMENT TITLE TYPE DOCUMENT INTERVAL Format 215 Unmatched listing As required Format 310 Test equipment inventory Monthly Format 350 Test equipment inventory in sub- custodian order Monthly Format 804 Recall schedule for on-site equipment Monthly/Quarterly Replenishment cards Preprinted METER card As required Blank METER cards Initial issue

1.4.5 Test Equipment References Several publications that contain information concerning test equipment are required to be maintained aboard ship by type commander instructions. These requirements are usually found in the inspection checkoff list. Other publications, while not required by directive, are necessary to you as reference and study material so you will be able to administer an effective test equipment program. Technicians should become familiar with the publications/directives listed in appendix II of this module.

p. 25

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-15 UNCLASSIFIED 1.5 INTRODUCTION TO TROUBLESHOOTING Our military forces increasingly rely on electrical and electronic equipment to help perform their mission. The effectiveness of our tactical forces depends on many types of electronic systems, such as communications systems, detection systems, and fire control systems. The reliability of such equipment is determined by many factors; however, the primary factors are the quality of the equipment in use, the availability of spare parts, and the ability of maintenance personnel to perform adequate maintenance.

Maintenance is work done to correct, reduce, or counteract wear, failure, and damage to equipment. Maintenance of electrical and electronic equipment is divided into two main categories: PREVENTIVE (routine) and CORRECTIVE maintenance. Preventive maintenance consists of mechanical, electrical, and electronic checks to determine whether equipment is operating properly. It also consists of visual inspections of cabling and equipment for damage and to determine if lubrication is needed. Corrective maintenance isolates equipment failure by means of test techniques and practices; it also replaces defective parts and realigns or readjusts equipment to bring it back to proper performance.

Q-9. What are the two main categories of maintenance? Q-10. What type of maintenance involves isolating equipment troubles and replacing defective parts? Testing and troubleshooting are the areas of maintenance that require the greatest technical skill. Testing procedures are referred to as measurements, tests, and checks. The definitions of these terms often overlap, depending on their use and the results obtained. For example, a power measurement and a frequency check could constitute a test of the operation of the same radio transmitter.

Troubleshooting is a term which we in the electronics field use daily. But what does it mean? Troubleshooting is sometimes thought to be the simple repair of a piece of equipment when it fails to function properly. This, however, is only part of the picture. In addition to repair, you, as a troubleshooter, must be able to evaluate equipment performance. You evaluate performance by comparing your knowledge of how the equipment should operate with the way it is actually performing. You must evaluate equipment both before and after repairs are accomplished.

p. 26

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-16 UNCLASSIFIED Equipment performance data, along with other general information for various electronic equipments, is available to help you in making comparisons. This information is provided in performance standards books for each piece of equipment. It illustrates what a particular waveform should look like at a given test point or what amplitude a voltage should be, and so forth. This data aids you in making intelligent comparisons of current and baseline operating characteristics for the specific equipment assigned to you for maintenance. ("Baseline" refers to the initial operating conditions of the equipment on installation or after overhaul when it is operating according to design.)

Remember, maintenance refers to all actions you perform on equipment to retain it in a serviceable condition or to restore it to proper operation. This involves inspecting, testing, servicing, repairing, rebuilding, and so forth. Proper maintenance can be performed only by trained personnel who are thoroughly familiar with the equipment. This familiarity requires a thorough knowledge of the theory of operation of the equipment.

A logical and systematic approach to troubleshooting is of the utmost importance in your performance of electronics maintenance. Many hours have been lost because of time- consuming "hit-or-miss" (often referred to as "easter-egging") methods of troubleshooting.

1.5.1 General Test Equipment Information In any maintenance training program, one of your most important tasks is to learn the use of test equipment in all types of maintenance work. To be effective in maintenance work, you must become familiar not only with the common types of measuring instruments, but also with the more specialized equipment. Some examples of common types of typical measuring instruments are the ammeter, voltmeter, and ohmmeter; examples of specialized test equipment are the spectrum analyzer, dual-trace oscilloscope, and power and frequency meters.

1.5.2 Test Equipment Safety Precautions The electrical measuring instruments included in test equipment are delicately constructed and require certain handling precautions to prevent damage and to ensure accurate readings. In addition, to prevent injury to personnel, you must observe precautions while using test equipment. You can find a list of applicable instructions in appendix II of this module.

p. 27

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-17 UNCLASSIFIED 1.5.2.1 Instrument Precautions To prevent damage to electrical measuring instruments, you should observe the precautions relating to three hazards: mechanical shock, exposure to magnetic fields, and excessive current flow.

MECHANICAL SHOCK - Instruments contain permanent magnets, meters, and other components that are sensitive to shock. Heavy vibrations or severe shock can cause these instruments to lose their calibration accuracy.

EXPOSURE TO STRONG MAGNETIC FIELDS - Strong magnetic fields may permanently impair the accuracy of a test instrument. These fields may impress permanent magnetic effects on permanent magnets, moving-coil instruments, iron parts of moving-iron instruments, or in the magnetic materials used to shield instruments.

EXCESSIVE CURRENT FLOW - This includes various precautions, depending on the type of instrument. When in doubt, use the maximum range scale on the first measurement and shift to lower range scales only after you verify that the reading can be made on a lower range. If possible, connections should be made while the circuit is de- energized. All connections should be checked to ensure that the instrument will not be overloaded before the circuit is reenergized.

1.5.2.2 Other Instrument Precautions Precautions to be observed to prevent instrument damage include the following:

• Keep in mind that the coils of wattmeters, frequency meters, and power meters may be carrying large quantities of current even when the meter pointer is on scale. • Never open secondaries of current transformers when the primary is energized. • Never short-circuit secondaries of potential transformers the primary is energized. • Never leave an instrument connected with its pointer off-scale or deflected in the wrong direction. • Ensure that meters in motor circuits can handle the motor starting current. This may be as high as six to eight times the normal running current. • Never attempt to measure the internal resistance of a meter movement with an ohmmeter since the movement may be damaged by the current output from the ohmmeter. • Never advance the intensity control of an oscilloscope to a position that causes an excessively bright spot on the screen; never permit a sharply focused spot to remain stationary for any period of time. This results in burn spots on the face of the cathode-ray tube (CRT).

p. 28

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-18 UNCLASSIFIED • In checking electron tubes with a tube tester that has a separate "short test," always make the short test first. If the tube is shorted, no further test should be made. • Before measuring resistance, always discharge any capacitors in the circuit to be tested. Note and record any points not having bleeder resistors or discharge paths for capacitors. • Always disconnect voltmeters from field generating or other highly inductive circuits before you open the circuit.

Q-11. Which quantity (voltage or current) determines the intensity of an electrical shock? Situations can arise during the use of test equipment that are extremely dangerous to personnel. For example, you may have an oscilloscope plugged into one receptacle, an electronic meter plugged into another, and a soldering iron in still another. Also, you may be using an extension cord for some equipments and not others or may be using other possible combinations. Some of the hazards presented by situations such as these include contact with live terminals or test leads. In addition, cords and test leads may be cross connected in such a manner that a potential difference exists between the metal cases of the instruments. This potential difference may cause serious or fatal shocks.

Test leads attached to test equipment should, if possible, extend from the back of the instruments away from the observer. If this is not possible, they should be clamped to the bench or table near the instruments.

At times, you may use instruments at locations where vibration is present, such as near a diesel engine. At such times, the instruments should be placed on pads of folded cloth, felt, or similar shock-absorbing material.

p. 29

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-19 UNCLASSIFIED 1.5.3 Working on Energized Circuits Insofar as is practical, you should NOT undertake repair work on energized circuits and equipment. However, it could become necessary, such as when you make adjustments on operating equipment. In such cases, obtain permission from your supervisor, then proceed with your work, but carefully observe the following safety precautions:

• DO NOT WORK ALONE. • Station an assistant near the main switch or circuit breaker so the equipment can be immediately de-energized in case of an emergency. • Someone qualified in first aid for electrical shock should be standing by during the entire operation. • Ensure that you have adequate lighting. You must be able to see clearly if you are to perform the job safely and properly. • Be sure that you are insulated from ground by an approved rubber mat or layers of dry canvas and/or wood. • Where practical, use only one hand, keeping the other either behind you or in your pocket. • If you expect voltage to exceed 150 volts, wear rubber gloves. • DO NOT work on any type of electrical apparatus when you are wearing wet clothing or if your hands are wet. • DO NOT wear loose or flapping clothing. • The use of thin-soled shoes and shoes with metal plates or hobnails is prohibited. • Flammable articles, such as celluloid cap visors, should not be worn. • Remove all rings, wristwatches, bracelets, and similar metal items before working on the equipment. Also ensure that your clothing does not contain exposed metal fasteners, such as zippers, snaps, buttons, and pins. • Do not tamper with interlock switches; that is, do not defeat their purpose by shorting them or blocking them open. • Ensure that equipment is properly grounded before energizing. • De-energize equipment before attaching alligator clips to any circuit. • Use only approved meters and other indicating devices to check for the presence of voltage.

p. 30

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-20 UNCLASSIFIED • Observe the following procedures when measuring voltages in excess of 300 volts: o Turn off the equipment power. o Short-circuit or ground the terminals of all components capable of retaining a charge. o Connect the meter leads to the points to be measured. o Remove any terminal grounds previously connected. o Turn on the power and observe the voltage reading. o Turn off the power. o Short circuit or ground all components capable of retaining a charge. o Disconnect the meter leads. • On all circuits where the voltage is in excess of 30 volts and where decks, bulkheads, or workbenches are made of metal, you should insulate yourself from accidental grounding by using approved insulating material. The insulating material should have the following qualities: o It should be dry, without holes, and should not contain conducting materials. o The voltage rating for which it is made should be clearly marked on the material. The proper material should be used so that adequate protection from the voltage can be supplied. o Dry wood may be used or, as an alternative, several layers of dry canvas, sheets of phenolic (resin or plastic) insulating material, or suitable rubber mats. o Care should be exercised to ensure that moisture, dust, metal chips, and so forth, which may collect on insulating material, are removed at once. Small deposits of such materials can become electrical hazards. o All insulating materials on machinery and in the area should be kept free of oil, grease, carbon dust, and so forth, since such deposits destroy insulation.

p. 31

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-21 UNCLASSIFIED 1.5.4 Safety Shorting Probe A representative shorting probe is shown in figure 1-4. An approved shorting probe is shown in NAVSEA 0967-LP-000-0100, EIMB, General, Section 3.

CAUTION

Capacitors and cathode-ray tubes may retain their charge for a considerable period of time after having been disconnected from the power source.

Always assume there is a voltage present when working with circuits having high capacitance, even when the circuit has been disconnected from its power source.

An approved type of shorting probe should be used to discharge capacitors and cathode-ray tubes individually.

When using the safety shorting probe, always be sure to first connect the test clip to a good ground (if necessary, scrape the paint off the grounding metal to make a good contact). Then hold the safety shorting probe by the insulated handle and touch the probe end of the shorting rod to the point to be shorted out. The probe end is fashioned so that it can be hooked over the part or terminal to provide a constant connection by the weight of the handle alone. Always take care not to touch any of the metal parts of the safety shorting probe while touching the probe to the exposed "hot" terminal. It pays to be safe; use the safety shorting probe with care.

Some equipment is provided with walk-around shorting devices, such as fixed grounding studs or permanently attached grounding rods. When that is the case, the walk-around shorting devices should be used rather than the safety shorting probe.

Figure 1-4 Representative safety shorting probe

p. 32

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-22 UNCLASSIFIED Q-12. What tool is used to de-energize capacitors in a circuit that has been disconnected from its power source? 1.5.5 Working on De-Energized Circuits When any electronic equipment is to be repaired or overhauled, certain general safety precautions should be observed. They are as follows:

• Remember that electrical and electronic circuits often have more than one source of power. Take time to study the schematics or wiring diagrams of the entire system to ensure that all sources of power have been disconnected • If pertinent, inform the remote station regarding the circuit on which work will be performed. • Use one hand when turning switches on or off. • Safety devices, such as interlocks, overload relays, and fuses, should never be altered or disconnected except for replacement. In addition, they should never be changed or modified in any way without specific authorization. • Fuses should be removed and replaced only after the circuit has been de- energized. When a fuse "blows," the replacement should be of the same type and have the same current and voltage ratings. A fuse puller should be used to remove and replace cartridge fuses. • All circuit breakers and switches from which power could possibly be supplied should be secured (locked if possible) in the OPEN or OFF (safe) position and danger tagged in accordance with procedures in the Standard Organization and Regulations of the U.S. Navy, OPNAVINST 3120.32. • After the work has been completed, the tag (or tags) should be removed only by the same person who signed it (them) when the work began. • Keep clothing, hands, and feet dry if at all possible. When you must work in wet or damp locations, place a rubber mat or other nonconductive material on top of a dry, wooden platform or stool; then use the platform or stool to sit and stand on. Use insulated tools and insulated flashlights of the molded type when you are required to work on exposed parts.

p. 33

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-23 UNCLASSIFIED 1.5.6 Grounding of Power Tools and Equipment The possibility of electrical shock can be reduced by ensuring that all motor and generator frames, metal bases, and other structural parts of electrical and electronic equipment are at ground potential.

Normally, on steel-hull vessels, such grounds are inherently provided because the metal cases or frames of the equipment are in contact with one another and with the metal structure of the vessel. In some instances where such inherent grounding is not provided by the mounting arrangements, such as equipment supported on shock mounts, suitable ground connections must be provided.

The grounding wire used for this purpose is generally made of flexible material (copper or aluminum) that provides sufficient current-carrying capacity to ensure an effective ground. In this manner, equipment cases and frames that are not intended to be above ground potential are effectively grounded; also, the possibility of electrical shock to personnel coming in contact with metal parts of the equipment is minimized. The secondary purpose of grounding equipment is to improve the operation and continuity of service of all equipments.

Paint, grease, or other foreign matter can interfere with the positive metal-to-metal contact at the ground connection point. Therefore, all bonding surfaces (connection points or metallic junctions) must be securely fastened and free of such matter. In all instances where equipment grounding is provided, certain general precautions and preventive maintenance measures must be taken. A few of these precautions are listed below:

• Periodically clean all strap-and-clamp connectors to ensure that all direct metal- to-metal contacts are free from foreign matter. • Check all mounting hardware for mechanical failure or loose connections. • Replace any faulty, rusted, or otherwise unfit grounding strap, clamp, connection, or component between the equipment and the ground to the ship’s hull. • When replacing a part of the ground connection, make certain that the metallic contact surfaces are clean and that electrical continuity is re-established. • After completing the foregoing steps, recheck to be sure that the connection is securely fastened with the correct mounting hardware. Paint the ground strap and hardware in accordance with current procedures.

p. 34

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-24 UNCLASSIFIED Because of the electrical shock hazards that could be encountered aboard ship, plugs and convenience outlets for use with portable equipment and power tools normally are standard three-prong type. Both plugs and outlets are keyed so that the plug must be in the correct position before it can be inserted into the receptacle. To ensure that the safety factors incorporated in these devices are in serviceable condition and are safe for use, you must perform the following precautions and inspections:

• Inspect the pins of the plug to see that they are firmly in place and are not bent or damaged. • Check the wiring terminals and connections of the plug. Loose connections and frayed wires on the plug surface must be corrected and any foreign matter removed before the plug is inserted into the receptacle. • Use a meter to ensure that the ground pin has a resistance of less than 1 ohm equipment ground. • Do not attempt to insert a grounded-type plug into a grounded receptacle without first aligning the plug properly.

CAUTION

Never use a power tool or a piece of portable test equipment unless you are absolutely sure that it is equipped with a properly grounded conductor.

p. 35

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-25 UNCLASSIFIED 1.6 BASIC MEASUREMENTS Electronic measurements involve the fundamental electrical quantities of voltage and current and the inherent characteristics of resistance, capacitance, and inductance. In circuits being tested, voltage and current are dependent upon resistance, capacitance, and inductance for their distribution; therefore, voltage and current measurements are valuable aids in determining circuit component conditions and in the evaluation of symptoms. Practically any reading obtained from the use of test equipment will depend on these basic measured quantities of resistance, capacitance, and inductance.

1.6.1 Voltage and Current Measurements Voltage measurements may be made as part of either preventive or corrective maintenance. These measurements are made using a voltmeter. When compared with voltage charts, these measurements are a valuable aid in locating a trouble quickly and easily. However, if the sensitivity of the test voltmeter differs from that of the voltmeter used in preparing the chart, the voltage measurements must be evaluated before the true circuit conditions can be determined. (Sensitivity in voltmeters was discussed in NEETS, Module 3, Introduction to Circuit Protection, Control, and Measurement.)

Since many of the troubles you find in equipments and systems are the result of abnormal voltages, voltage measurements are a valuable aid in locating trouble. You can measure voltage with a voltmeter without interrupting circuit operation.

Point-to-point voltage measurement charts, usually found in equipment technical manuals, contain the normal operating voltages found in the various stages of the equipment. These voltages are usually measured between indicated points and ground unless otherwise stated. When you begin recording voltage measurements, it is a smart and safe practice to set the voltmeter on the highest range before measuring. This ensures that excessive voltages existing in the circuit will not cause overloading of the meter.

Q-13. On what range should you set the voltmeter prior to taking a voltage measurement? To increase accuracy, you should then set the voltmeter to the appropriate range for the proper comparison with the expected voltage in the voltage charts. When checking voltages, remember that a voltage reading can be obtained across a resistance, even if that resistance is open. The resistance of the meter itself forms a circuit resistance when the meter probes are placed across the open resistance. Therefore, the voltage across the component may appear to be normal or near-normal as you read the meter, but may actually be abnormal when the meter is disconnected from the circuit.

p. 36

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-26 UNCLASSIFIED If the internal resistance of the voltmeter is approximately the same value as the resistance being tested, it will indicate a considerably lower voltage than the actual voltage present when the meter is removed from the circuit. The sensitivity (in ohms per volt) of the voltmeter used to prepare the voltage charts is provided on those charts. If a meter of similar sensitivity is available, you should use it to reduce the effects of loading.

The following precautions are general safety measures that apply to the measurement of voltages. Remember that nearly all voltages are dangerous and have often proved fatal to careless technicians. When measuring voltages, be sure to observe the following precautions:

• Set test equipment to the HIGHEST range. • Make sure safety observer knows where to secure power for the equipment under test. • Connect the ground lead of the voltmeter first. • Use only one hand to take measurements (when possible), and put the other hand in your pocket or behind your back. • If the voltage to be measured is less than 300 volts, place the end of the test probe on the point to be tested; use the polarity switch to select positive or negative readings. • If the voltage to be measured is more than 300 volts, proceed as follows: o Shut off circuit power. o Discharge all filter capacitors with a shorting probe. o Temporarily ground the point to be measured. o Connect (clip on) the proper test lead to the high-voltage point. o Move away from the voltmeter. o Turn on circuit power and read the voltmeter. o Turn off circuit power. o Discharge all capacitors before disconnecting the meter.

Q-14. When taking a voltage measurement, which lead of the voltmeter should you connect to the circuit first? Current measurements are not often taken in the course of preventive maintenance or testing. This is because the ammeter (or other current-measuring instrument) must become an actual part of the equipment being tested. The circuit must be opened (desoldered) to connect the ammeter in series with the circuit being tested. Usually, you can take a voltage measurement and use this factor to calculate the circuit current by applying Ohm’s law.

Q-15. Is an ammeter connected in series or in parallel with the circuit under test?

p. 37

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-27 UNCLASSIFIED 1.6.2 Resistance Measurements Resistance measurements are a valuable aid to you in locating defective circuits and components during corrective maintenance. Maintenance handbooks for the equipment can often be used to help you take these measurements. These handbooks often contain resistance charts that are referenced to accessible test points within the equipment. Without these charts, taking resistance measurements in a complex circuit is a slow process. The process is slow because one side of the circuit component must often be desoldered to get a true resistance measurement. However, resistance tolerances vary so widely that approximate resistance readings are adequate for most jobs.

Once the most accessible test point is found, an ohmmeter is usually used to take the resistance measurement. Because of the degree of accuracy needed when an ohmmeter is used, proper calibration and understanding of the meter scales is a must. (Topic 2 of this module will discuss these requirements in detail.) When using an ohmmeter, you must observe the following precautions:

• The circuit being tested must be completely de-energized. • Any meters or transistors which can be damaged by the ohmmeter current must be removed before any measurement is made.

Q-16. What must be done to a circuit before you can use an ohmmeter for testing? 1.6.3 Capacitance Measurements Capacitance measurements are usually taken with a capacitance meter. Capacitance tolerances vary even more widely than resistance tolerances. Capacitance tolerances depend on the type of capacitor, the value of capacitance, and the voltage rating. The actual measurement of capacitance is very simple; however, you must make the important decision of whether to reject or to continue to use the capacitor after it has been tested.

The POWER FACTOR of a capacitor is important because it is an indication of the various losses of a capacitor. Power losses can be traced to the dielectric, such as current leakage and dielectric absorption. Current leakage is of considerable importance, especially in electrolytic capacitors.

Q-17. What is the term used to refer to the losses which can be traced to the dielectric of a capacitor?

p. 38

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-28 UNCLASSIFIED 1.6.4 Inductance Measurements Inductance measurements are seldom required in the course of troubleshooting. However, inductance measurements are useful in some cases; therefore, bridges (discussed in the next section) are available for making this test. You will find that many capacitance test sets can be used to measure inductance. Most capacitance test sets are furnished with inductance conversion charts if the test equipment scale is not calibrated to read the value of inductance directly.

1.6.5 Capacitance, Inductance, and Resistance Bridges You can measure capacitance, inductance, and resistance for precise accuracy by using ac bridges. These bridges are composed of capacitors, inductors, and resistors in a wide variety of combinations. These bridges are operated on the principle of a dc bridge called a WHEATSTONE BRIDGE.

1.6.5.1 Wheatstone Bridge The Wheatstone bridge is widely used for precision measurements of resistance. The circuit diagram for a Wheatstone bridge is shown in figure 1-5. Resistors R1, R2, and R3 are precision, variable resistors. The value of Rx is an unknown value of resistance that must be determined. After the bridge has been properly balanced (galvanometer G reads zero), the unknown resistance may be determined by means of a simple formula. The galvanometer (an instrument that measures small amounts of current) is inserted across terminals b and d to indicate the condition of balance. When the bridge is properly balanced, no difference in potential exists across terminals b and d; when switch S2 is closed, the galvanometer reading is zero.

p. 39

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-29 UNCLASSIFIED

The operation of the bridge is explained in a few logical steps. When the battery switch S1 is closed, electrons flow from the negative terminal of the battery to point a. Here the current divides as it would in any parallel circuit. Part of it passes through R1 and R2; the remainder passes through R3 and Rx. The two currents, I 1 and I2, unite at point c and return to the positive terminal of the battery. The value of I1 depends on the sum of resistance R1 and R2, and the value of I2 depends on the sum of resistances R3 and Rx. In each case, according to Ohm’s law, the current is inversely proportional to the resistance.

Figure 1-5 Wheatstone bridge

p. 40

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-30 UNCLASSIFIED R1, R2, and R3 are adjusted so that when S1 is closed, no current flows through G. When the galvanometer shows no deflection, there is no difference of potential between points b and d. All of I1 follows the a b c path and all I2 follows the a b c path. This means that a voltage drop E1 (across R1 between points a and b) is the same as voltage drop E3 (across R3 between points a and d). Similarly, the voltage drops across R2 and Rx (E2 and Ex) are also equal. Expressed algebraically,

E1 = E3

I1R1 = I2R3

and

E2 = Ex

I1R2 = I2Rx

With this information, we can figure the value of the unknown resistor R x. Divide the voltage drops across R1 and R3 by their respective voltage drops across R2 and Rx as follows:

I1R1 I1R2 = I2R3 I2Rx

We can simplify this equation:

R1 R2 = R3 Rx

then we multiply both sides of the expression by Rx to separate it:

Rx = R2R3 Rx

For example, in figure 1-5, we know that R1 is 60 ohms, R2 is 100 ohms, and R3 is 200 ohms. To find the value of R x, we can use our formula as follows:

Rx = R2R3 R1

Rx = 100 × 200 60

p. 41

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-31 UNCLASSIFIED Rx = 20,000 60

Rx = 333.33 ohms

1.6.5.2 Use of ac Bridges A wide variety of ac bridge circuits (such as the Wheatstone) may be used for the precision measurement of ac resistance, capacitance, and inductance. Let’s look at ac bridges in terms of functions they perform.

RESISTANCE BRIDGE - An ac signal generator, as shown in figure 1-6, is used as the source of voltage. Current from the generator passes through resistors R1 and R2, which are known as the ratio arms, and through Rs and Rx. Again, Rx is known as resistance. Rs has a standard value and replaces R3 in figure 1-6. When the voltage drops across R2 and Rs are equal, the voltage drops across R2 and Rx are also equal; no difference of potential exists across the meter and no current flows through it. As we discovered with the Wheatstone bridge, when no voltage appears across the meter, the following ratio is true:

Figure 1-6 Resistance bridge (ac)

p. 42

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-32 UNCLASSIFIED For example, if in figure 1-6 we know that R1 is 20 ohms, R2 is 40 ohms, and Rs is 60 ohms, we can find the value of Rx using our formula as follows:

Rx = R2Rs R1

Rx = 40 × 60 20

Rx = 2,400 20

Rx = 120 ohms

With the ac signal applied to the bridge, R1 and R2 are varied until a zero reading is seen on the meter. Zero deflection indicates that the bridge is balanced. (NOTE: In actual practice, the variables are adjusted for a minimum reading since the phase difference between the two legs will not always allow a zero reading.)

CAPACITANCE BRIDGE - Because current varies inversely with resistance and directly with capacitance, an inverse proportion exists between the four arms of the bridge in figure 1-7; the right side of our expression is inverted from the resistance bridge expression as follows:

R1 R2 = Cx Cs

Cx = R1Cs R2

p. 43

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-33 UNCLASSIFIED

Q-18. What effect does an increase in capacitance have on a capacitor’s opposition to current flow? Because R1 and R2 are expressed in the same units, the equation R1/R2 becomes a simple multiplication factor. This equation provides a numerical value for Cx and will be in the same units as Cs (farad, microfarad, and so forth).

Similarly, the following resistance ratio exists between the four arms of the bridge, just as in the resistance bridge expression discussed earlier:

R1 R2 = Rs Rx

or

Rx = R2Rs R1

Figure 1-7 Capacitance bridge

p. 44

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-34 UNCLASSIFIED Thus, both the unknown resistance and capacitance, Rx and Cx, can be estimated in terms of known resistance R1, R2, Rs, and known capacitance Cs.

In figure 1-7, for example, we know that R1 is 20 ohms, R2 is 40 ohms, Rs is 60 ohms, and Cs is 10 microfarads. We can find the values of Cx and Rx by using the respective formulas as follows:

Cx = R1Cs R2

Cx = 20 × 10 40

Cx = 200 40

Cx = 5 microfarads

and

Rx = R2Rs R1

Rx = 40 × 60 20

Rx = 2,400 20

Rx = 120 ohms

Q-19. When a bridge is used to measure resistance, what is the value of R x if R1 equals 80 ohms, R2 equals 120 ohms, and R3 equals 280 ohms?

p. 45

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-35 UNCLASSIFIED INDUCTANCE BRIDGE - The value of the unknown inductance Lx may be determined by means of the simple bridge circuit shown in figure 1-8. Ratio arms R1 and R2 are accurately calibrated resistors. Ls is a standard inductor with a known inductance; Rs is the known resistance, and Rx represents the resistance of the unknown inductor.

The ac signal is applied to the bridge, and variable resistors R1 and R2 are adjusted for a minimum or zero deflection of the meter, indicating a condition of balance. When the bridge is balanced, the following formulas may be used to find Lx.

(NOTE: The right side of this expression is NOT inverse as it was in the capacitance bridge.)

R1 R2 = Ls Lx

R2Ls R1 = Lx

Figure 1-8 Inductance bridge

p. 46

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-36 UNCLASSIFIED and

R1 R2 = Rs Rx

or

Rx = R2Rs R1

In figure 1-8, for example, the values of R1, R2, and Rs are 20, 40, and 60 ohms, respectively. The value of Ls is 10 millihenries. We can find the values of Rx and Lx by using their respective formulas as follows:

Lx = R2Ls R1

Lx = 40 × 10 20

Lx = 400 20

Lx = 20 millihenries

and

Rx = R2Rs R1

Rx = 40 × 60 20

Rx = 2,400 20

Rx = 120 ohms

Thus, both the unknown resistance and inductance can be estimated in terms of the known values for R1, R2, R s, and Ls.

Q-20. When an unknown capacitance is tested with a bridge, what is the value of C x if R1 equals 70 ohms, R2 equals 150 ohms, and Cs equals 550 microfarads?

p. 47

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-37 UNCLASSIFIED 1.7 SUMMARY The important points of this chapter are summarized in the following paragraphs:

The JETDS SYSTEM is jointly used by all branches of the military to identify equipments by a system of standardized nomenclatures.

GPETE is test equipment that has the capability, without modifications, to generate, modify, or measure a range of parameters of electronic functions required to test two or more equipments or systems of basically different design. All GPETE are listed in Standard General Purpose Electronic Test Equipment, MIL-STD-1364 (series).

SPETE is test equipment that is specifically designed to generate, modify, or measure a range of parameters of electronic functions of a specific or peculiar nature required to test a single equipment or system.

The SHIP CONFIGURATION AND LOGISTICS INFORMATION SYSTEM (SCLSIS) program is designed to keep track of equipment configuration changes in the fleet.

The SCLSIS program has two basic elements, VALIDATION and INVENTORY UPDATING.

The CALIBRATION STATUS of any items of test equipment can be determined by the information recorded on the calibration label or tag located on the equipment.

The CALIBRATED label, with black lettering on a white background, indicates the instrument to which it is attached is within tolerance on all scales.

The CALIBRATED—REFER TO REPORT label, with red lettering on a white background, is used when actual measurement values must be known to use the instrument.

The SPECIAL CALIBRATION label, with black lettering on a yellow background, is used when some unusual or special condition in the calibration should be drawn to your attention.

The USER CALIBRATION label indicates that you should calibrate the test and measuring instrument instead of sending the instrument to a calibration facility.

The INACTIVE—CALIBRATE BEFORE USE label is used when a piece of test equipment due for recalibration will not be used for some time in the future.

p. 48

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-38 UNCLASSIFIED The CALIBRATION NOT REQUIRED label is used on test instruments listed in the Metrology Requirements List (METRL) as not requiring calibration.

The REJECTED label is attached to a test instrument that fails to meet the acceptance criteria during calibration and cannot be repaired.

The CALIBRATION VOID IF SEAL BROKEN label is placed over readily accessible adjustments to prevent tampering by the user when such tampering could affect the calibration.

The MEASURE system is designed to standardize the recall and scheduling of test, measurement, and diagnostic equipment into calibration facilities and for the documentation of actions performed by the calibration facility.

MAINTENANCE is work done to correct, reduce, or counteract wear and damage to equipment.

PREVENTIVE MAINTENANCE consists of checks to determine weather equipment is functioning properly. It also consists of visual inspections of cabling and equipment for damage and to determine if lubrication is needed.

CORRECTIVE MAINTENANCE is used to isolate troubles by means of test techniques and practices that realign or readjust equipment or otherwise bring the equipment back up to proper performance.

SENSITIVITY of the voltmeter is always given on the voltage charts for a particular piece of equipment. You should always use a voltmeter of similar sensitivity to the equipment to diminish the effects of circuit loading.

CURRENT MEASUREMENTS are not often taken in the course of testing because the ammeter (or other current measuring device) must become an actual part of the equipment being tested. The circuit must be opened for necessary connection of the meter. Usually you can use a voltage measurement to calculate the circuit current by applying Ohm’s law.

You should observe the following PRECAUTIONS when using an ohmmeter:

1. The circuit being tested must be completely de-energized. 2. Any circuit components which can be damaged by ohmmeter current must be removed before any measurement is made.

The WHEATSTONE BRIDGE is used for precise measurement of resistance.

p. 49

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-39 UNCLASSIFIED The CAPACITANCE BRIDGE is used for measuring an unknown capacitance.

An INDUCTANCE BRIDGE is used to find the value of an unknown inductance

p. 50

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

1-40 UNCLASSIFIED ANSWERS TO QUESTIONS Q1. THROUGH Q20.

A-1. Joint Electronics Type Designation System (JETDS).

A-2. General-purpose electronic test equipment (GPETE) and special-purpose electronic test equipment (SPETE).

A-3. Special-purpose electronic test equipment.

A-4. Validation and updating.

A-5. CALIBRATED— REFER TO REPORT.

A-6. SPECIAL CALIBRATION label.

A-7. Maintenance personnel.

A-8. The Chief of Naval Operations.

A-9. Preventive and corrective maintenance.

A-10. Corrective maintenance.

A-11. Current.

A-12. Shorting probe.

A-13. Highest.

A-14. Ground.

A-15. In series.

A-16. It must be de-energized.

A-17. Power losses.

A-18. Opposition to current flow decreases.

A-19. 420 ohms.

A-20. 256 microfarads.

Chapter 3 Basic Meters

p. 87

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-1 UNCLASSIFIED 3 BASIC METERS LEARNING OBJECTIVES

Upon completing this chapter, you should be able to:

1. Describe the basic theory of the galvanometer. 2. Describe the basic theory of the D’Arsonval meter movement. 3. State the proper procedure for connecting an ammeter to a circuit. 4. Define ammeter sensitivity. 5. State the proper procedure for connecting a voltmeter to a circuit. 6. Describe possible effects on a circuit caused by the connection of a voltmeter. 7. Define voltmeter sensitivity. 8. Describe the internal operation of an ohmmeter with the use of a block diagram. 9. Describe the operating procedure for using a megohmmeter. 10. Describe the use of the electrodynamometer-type meter as a voltmeter, ammeter, and wattmeter. 11. Describe the factors that limit wattmeter capability. 12. Describe an open circuit, a ground, a short, and the tests used to check for these conditions.

3.1 INTRODUCTION When troubleshooting, testing, or repairing electronic equipment, you will use various meters and other types of test equipment to check for proper circuit voltages, currents, resistances, and to determine if the wiring is defective. You may be able to connect these test instruments to a circuit and take readings without knowing just how the instruments operate. However, to be a competent technician, you need to be able to do more than merely read a test instrument. You need a basic knowledge of how test instruments operate. This chapter discusses the operating principles of some of the test instruments you will use in equipment troubleshooting.

p. 88

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-2 UNCLASSIFIED 3.2 METERS The best and most expensive measuring instrument is of no use to you unless you know what you are measuring and what each reading indicates. Remember that the purpose of a meter is to measure quantities existing within a circuit. For this reason, when the meter is connected to the circuit, it must not change the condition of the circuit.

3.2.1 Meter Power Source Meters are either SELF-EXCITED or EXTERNALLY EXCITED. Self-excited meters operate from their own power sources. Externally excited meters get their power from the circuit to which they are connected. Most common meters (voltmeters, ammeters, and ohmmeters) that you use in your work operate on the electromagnetic principle. All measuring instruments must have some form of indicating device, usually a meter, to be of any use to you. The most basic indicating device used in instruments that measure current and voltage operates by using the interaction between the magnetic fields associated with current flow in the circuit. Before continuing, you might want to review the properties of magnetism and electromagnetism in NEETS, Module 1, Introduction to Matter, Energy, and Direct Current.

Q-1. What meters operate from their own power sources? 3.2.2 Basic Meter Movement A stationary, permanent-magnet, moving-coil meter is the basic meter movement used in most measuring instruments used for servicing electrical equipment. When current flows through the coil, a resulting magnetic field reacts with the magnetic field of the permanent magnet and causes the movable coil to rotate. The greater the intensity of current flow through the coil, the stronger the magnetic field produced; the stronger the magnetic field produced, the greater the rotation of the coil. The GALVANOMETER is an example of one type of stationary, permanent-magnet, moving-coil measuring instrument.

p. 89

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-3 UNCLASSIFIED 3.2.2.1 Galvanometer The galvanometer is used to measure very low currents, such as those in bridge circuits. In modified form, the galvanometer has the highest sensitivity of any of the various types of meters in use today. A simplified diagram of a galvanometer is shown in figure 3-1. It is different from other instruments used for the same purpose because its movable coil is suspended by means of metal ribbons instead of a shaft and jewel-bearing arrangement often used in other instruments.

The movable coil is wrapped around the aluminum frame of the galvanometer. The coil is suspended between the poles of the magnet by means of thin, flat ribbons of phosphor bronze. These ribbons provide a conduction path for the current between the circuit being tested and the movable coil. The ribbons allow the coil to twist in response to the interaction of the applied current through the coil and the magnetic field of the permanent magnet. They also provide the restoring force for the coil. Basically, the restoring force is that force necessary to return the movable frame to its resting position after a reading. The ribbons restrain or provide a counterforce to the magnetic force acting on the coil. When the driving force of the coil current is removed, the restoring force provided by the ribbons returns the coil to its zero position. Figure 3-1 Simplified galvanometer

p. 90

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-4 UNCLASSIFIED Q-2. What physical component of a galvanometer provides the restoring force for the coil? To determine the amount of current flow, we must have a means to indicate the amount of coil rotation. Either of two methods may be used: (1) the POINTER arrangement or (2) the LIGHT AND MIRROR arrangement.

Q-3. In a galvanometer, what two methods are used to indicate the amount of coil rotation? In the pointer arrangement, one end of the pointer is mechanically connected to the rotating coil; as the coil moves, the pointer also moves. The other end of the pointer moves across a graduated scale and indicates the amount of current flow. The overall simplicity of this arrangement is its main advantage. However, a disadvantage of this arrangement is that it introduces a mechanical coil balancing problem, especially if the pointer is long.

Q-4. What is the primary disadvantage of the pointer arrangement for indicating coil rotation? In the light and mirror arrangement, the use of a mirror and a beam of light simplifies the problem of coil balance. When this arrangement is used to measure the turning of the coil, a small mirror is mounted on the supporting ribbon, as shown in figure 3-1. An internal light source is directed to the mirror and then reflected to the scale of the meter. As the movable coil turns, so does the mirror. This causes the light reflection to move across the graduated scale of the meter. The movement of the reflection is proportional to the movement of the coil; therefore, the intensity of the current being measured by the meter is accurately indicated.

If the beam of light and mirror arrangement is used, the beam of light is swept to the right or left across a translucent screen (scale). The translucent screen is divided uniformly with the zero reading located at center scale. If the pointer arrangement is used, the pointer is moved in a horizontal plane to the right or left across a scale that is divided uniformly with the zero reading at the center. The direction in which the beam of light or the pointer moves depends on the direction (polarity) of current through the coil.

p. 91

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-5 UNCLASSIFIED 3.2.2.2 D’Arsonval Meter Movement Most dc instruments use meters based on some form of the D’Arsonval meter movement. In D’Arsonval-type meters, the length of the conductor and the strength of the field between the poles of the magnet are fixed. Therefore, any change in current causes a proportional change in the force acting on the coil. Figure 3-2 is a simplified diagram showing the principle of the D’Arsonval movement.

In the figure, only one turn of wire is shown; however, in an actual meter movement, many turns of fine wire would be used, each turn adding more effective length to the coil. The coil is wound on an aluminum frame (bobbin) to which the pointer is attached. Oppositely wound hairsprings (only one is shown in the figure) are also attached to the bobbin, one at either end. The circuit to the coil is completed through the hairsprings. In addition to serving as conductors, the hairsprings serve as the restoring force that returns the pointer to the zero position when no current flows.

Q-5. What component of the D’Arsonval meter movement completes the circuit for current flow to the coil? Figure 3-2 D'Arsonval meter movement

p. 92

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-6 UNCLASSIFIED COIL MOVEMENT - As we discussed previously, the deflecting (moving) force on the coil is proportional to the current flowing through the coil. This deflecting force tends to cause the coil to rotate against the restraining force of the hairsprings. When the deflecting force and the restraining force are equal, the coil and the pointer stop moving. As we have just stated, the deflecting force is proportional to the current in the coil, the angle (amount) of rotation is proportional to the deflecting force; therefore, the angle of rotation is proportional to the current through the coil. When current stops flowing through the coil, the deflecting force stops, and the restoring force of the springs returns the pointer to the zero position.

Q-6. What component supplies restoring force to the coil of the D’Arsonval meter movement? DIRECTION OF FORCE - The current through the single turn of wire is in the direction indicated in the figure (away from you on the right-hand side and toward you on the left-hand side). If we apply the right-hand motor rule, the direction of force is upward on the left-hand side and downward on the right-hand side; therefore, the direction of motion of the coil and pointer is clockwise. If the current were reversed in the wire, the direction of motion of the coil and pointer would be reversed. For a review of the right- hand rule for motors, refer to NEETS, Module 5, Introduction to Generators and Motors.

PRINCIPLE OF OPERATION - A more detailed view of the basic D'Arsonval movement, as it is used in ammeters and voltmeters, is shown in figure 3-3. The principle of operation is the same as that discussed in the simplified version. The iron core is rigidly supported between the pole pieces; it serves to concentrate the flux in the narrow space between the iron core and the pole piece. Current flows into one hairspring, through the coil, and out the other hairspring. The restoring forces of the spiral springs return the pointer to the normal zero position when the current through the coil is interrupted. Conductors connect the hairsprings with the outside terminals of the meter. If the instrument is not DAMPED to absorb the energy of the moving element, the pointer will oscillate (vibrate) for a period of time before coming to a stop in its final position. Damping is an energy-absorbing system that prevents this.

p. 93

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-7 UNCLASSIFIED

Figure 3-3 Detailed view of the basic D'Arsonval meter movement

p. 94

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-8 UNCLASSIFIED DAMPING - This is accomplished in many D'Arsonval movements by means of the motion of the aluminum bobbin on which the coil is wound. As the bobbin rotates in the magnetic field, an electromotive force is induced into it as it cuts through the lines of force. Induced currents flow in the bobbin in a direction opposite to the motion; this causes the bobbin to go beyond its final position only once before stopping. The overall sensitivity of the meter can be increased by the use of a lightweight rotating assembly (bobbin, coil, and pointer) and by the use of jewel bearings, as shown in figure 3-3.

POLE CONSTRUCTION - Note that the pole pieces in figures 3-2 and 3-3 have curved faces. You can see the advantage of this type of construction if you remember that lines of force enter and leave a magnetic field in the air gap at right angles to the coil, regardless of the angular position of the coil. Because of this type of construction, a more linear scale is possible than if the pole faces were flat.

Q-7. What advantage is gained by using pole pieces with curved faces in the D’Arsonval meter movement? 3.2.3 DC Ammeter The movable coil of the D'Arsonval meter movement we have been discussing up to now uses small-size wire in its windings. This small-size wire places limits on the amount of current that can be safely passed through the coil. Therefore, the basic D'Arsonval movement discussed can be used to indicate or measure only very small currents. Certain circuit changes must be made to the basic D'Arsonval meter movement for it to be practical in everyday use. To measure large currents, you must use a SHUNT with the meter.

3.2.3.1 Shunts A shunt is a physically large, low-resistance conductor connected in parallel (shunt) with the meter terminals. It is used to carry the majority of the load current. Such a shunt is designed with the correct amount of resistance so that only a small portion of the total current flows through the meter coil. The meter current is proportional to the total load current. If the shunt is of such a value that the meter is calibrated in milliamperes, the instrument is called a MILLIAMMETER. If the shunt has such a value that the meter must be calibrated in terms of amperes, it is called an AMMETER.

Q-8. What structurally large, low-resistance conductor is connected in parallel with the meter movement to prevent damage?

p. 95

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-9 UNCLASSIFIED SHUNT RESISTANCE - A single, standardized meter movement is normally used in all ammeters, no matter what the range is for a particular meter. For example, meters with working ranges of 0 to 10 amperes, 0 to 5 amperes, or 0 to 1 ampere all use the same meter movement. The various ranges are achieved through the use of different values of shunt resistance with the same meter movement. The designer of the ammeter simply calculates the correct shunt resistance required to extend the range of the meter movement to measure any desired value of current. This shunt is then connected across the meter terminals. Shunts may be located inside the meter case (internal shunts) with the proper switching arrangements for changing them. They may also be located outside the meter case (external shunts) with the necessary leads to connect them to the meter.

EXTERNAL SHUNTS - An external-shunt circuit is shown in figure 3-4, view A. Typical external shunts are shown in view B. View C shows a meter movement mounted within the case. The case provides protection against breakage, magnetic shielding in some cases, and portability.

Figure 3-4 Dc ammeter using the D'Arsonval movement with external shunts

p. 96

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-10 UNCLASSIFIED SHUNT CONSTRUCTION - The shunt strips (view B of figure 3-4) are usually made of the alloy Manganin. Manganin has a temperature coefficient of almost zero. The zero- temperature coefficient property is desirable because of the heavy currents that often flow through shunts producing heat. A zero-temperature coefficient material is not affected by this heat; therefore, it remains stable in temperature. Most other materials increase their resistance as they are heated. If shunts were made of these materials, they would carry less current. More and more current would flow through the meter movement, and the chances of damage would increase. Using shunts constructed with zero-temperature coefficient materials eliminates this problem.

Q-9. What type of temperature coefficient material does not produce increased heat in response to increased current flow? The ends of the shunt strips are embedded in heavy copper blocks. The blocks are attached to the meter coil leads and the line terminals. To ensure accurate readings, you should not interchangeably use the meter leads for a particular ammeter with those for a meter of a different range. Slight changes in lead length and size may vary the resistance of the meter circuit. If this happens, current will also change and cause incorrect meter readings. External shunts are generally used where currents greater than 50 amperes must be measured.

SHUNT SELECTION - When using an external-shunt ammeter, you should select a suitable shunt so that the scale deflection can be easily read. For example, if the scale has 150 divisions and the load current you want to measure is known to be between 50 and 100 amperes, a 150-ampere shunt would be the correct choice. Under these conditions, each division of the scale represents 1 ampere. In other words, a full-scale deflection of the pointer would rest on the 150th division mark, indicating that 150 amperes of load current is flowing. At half-scale deflection, the pointer would rest on the 75th division mark, indicating that 75 amperes of load current is flowing.

A shunt having exactly the same current rating as the expected normal load current should never be selected. If you were to select such a shunt, higher than normal load currents could possibly drive the pointer off scale and damage the meter movement. A good choice of shunt values will place the indicating needle somewhere near the midscale indication when the load current you are reading is normal. For example, assume that the meter scale is divided into 100 equal divisions and you want to measure a current of 60 amperes. The shunt to use would be a 100-ampere shunt. This would make each division of the scale equal to 1 ampere. The meter indication would fall on the 60th division showing that 60 amperes of load current is flowing. Therefore, an allowance (40 amperes) remains for unexpected surge currents.

p. 97

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-11 UNCLASSIFIED Q-10. A good choice of shunt resistance will place the indicating pointer near what part of the meter scale with a normal load? INTERNAL SHUNTS FOR METERS IN THE 0- TO 50-AMPERE RANGE - When measuring current ranges below 50 amperes, you will most often use internal shunts (Rshunt). In this way, you can easily change the range of the meter by means of a switching arrangement. A switch will select the correct internal shunt with the necessary current rating and resistance. Before you can calculate the required resistance of the shunt for each range, the total resistance of the meter movement must be known. For example, suppose you desire to use a 100-microampere D'Arsonval meter with an internal coil resistance of 100 ohms to measure line currents up to 1 ampere. The meter will deflect to its full-scale position when the current through the deflection coil is 100 microamperes.

Since the coil resistance is 100 ohms, you can calculate the coil's voltage (E coil) by using Ohm's law, as follows:

Ecoil = I × R

= 0.0001 amperes × 100 ohms

= 0.01 volt

When the pointer is deflected to full scale, 100 microamperes of current flows through the coil and 0.01 volt drops across it. Remember, 100 microamperes is the maximum safe current for this meter movement. Exceeding this value will damage the meter. The shunt must carry any additional load current.

The meter coil has a 0.01 volt drop across it, and, because the shunt and coil are in parallel, the shunt also has a voltage drop of 0.01 volt. The current that flows through the shunt is the difference between the full-scale meter current and the line current being fed into the shunt. In this case, meter current is 100 microamperes. Full-scale deflection is desired only when the total current is 1 ampere. Therefore, the shunt current must equal 1 ampere minus 100 microamperes, or 0.9999 ampere. Ohm's law is again used to provide the approximate value of required shunt resistance (R shunt), as follows:

Rshunt = E I

0.01 volt 0.999 ampere

= 0.01 ohm

p. 98

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-12 UNCLASSIFIED To increase the range of the 100-microampere meter to 1 ampere (full-scale deflection), place a 0.01-ohm shunt in parallel with the meter movement.

You can convert the 100-microampere instrument to a 10-ampere meter by using a proper shunt. The voltage drop for a full-scale deflection is still 0.01 volt across the coil and the shunt. The meter current is still 100 microamperes. The shunt current must therefore be 9.9999 amperes under full-scale deflection. Again, this is an approximate figure found by the application of Ohm’s law.

You can also convert the same instrument to a 50-ampere meter by using the proper shunt resistance, as follows:

Rshunt = E I

0.01 volt 49.999 amperes

= 0.0002 ohm

INTERNAL SHUNTS FOR METERS IN THE MILLIAMPERE RANGE - The above method of computing the shunt resistance is satisfactory in most cases; however, it can only be used when the line current is in the ampere range and the meter current is relatively small compared to the load current. In such cases, you can use an approximate value of resistance for the shunt, as was done above. However, when the line current is in the milliampere range and the coil current becomes an appreciable percentage of the line current, a more accurate calculation must be made. For example, suppose you desire to use a meter movement that has a full-scale deflection of 1 milliampere and a coil resistance of 50 ohms to measure currents up to 10 milliamperes. Using Ohm's law, you can figure the voltage (E coil) across the meter coil (and the shunt) at full-scale deflection, as follows:

E coil = I × R

= 0.001 amperes × 50 ohms

= 50 millivolts

p. 99

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-13 UNCLASSIFIED The current that flows through the shunt (Ishunt) is the difference between the line current and the meter current, as figured below:

Ishunt = Itotal − Imeter

= 10mA − 1mA

= 9 mA or 0.009 ampere

The shunt resistance (Rshunt) may then be figured, as follows:

Rshunt = E I

0.05 volt 0.009 amperes

= 5.55 ohms

Notice that, in this case, the exact value of shunt resistance has been used rather than an approximation.

The formula for determining the resistance of the shunt is given by Rs = I m /Is times Rm, where Rs is the shunt resistance in ohms; Im is the meter current at full-scale deflection; Is is the shunt current at full-scale deflection; and Rm is the resistance of the meter coil. If the values given in the previous example are used in this equation, it will yield 5.55 ohms, the value previously calculated.

SWITCHING SHUNT VALUES - Various values of shunt resistance can be used, by means of a suitable switching arrangement, to increase the number of current ranges that can be covered by the meter. Two switching arrangements are shown in figure 3-5. View A is the simpler of the two arrangements when a number of shunts are used to calculate the values of the shunt resistors. However, it has two disadvantages:

p. 100

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-14 UNCLASSIFIED

1. When the switch is moved from one shunt resistor to another, the shunt is momentarily removed from the meter. The line current then flows through the meter coil. Even a momentary surge of current could easily damage the coil.

2. The contact resistance (resistance between the blades of the switch when they are in contact) is in series with the shunt, but not with the meter coil. In shunts that must pass high currents, this contact resistance becomes an appreciable part of the total shunt resistance. Because the contact resistance is of a variable nature, the ammeter indication may not be accurate.

The generally preferred method of range switching is shown in (figure 3-5, view B). Although only two ranges are shown, as many ranges as needed can be used. In this type of circuit, the contact resistance of the range-selector switch is external to the shunt and meter in each range position. The contact resistance in this case has no effect on the accuracy of the current measurement.

Figure 3-5 Ways of connecting internal meter shunts

p. 101

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-15 UNCLASSIFIED 3.2.3.2 Ammeter Connections When you are using ammeters, a primary rule of safety is that such current-measuring instruments must always be connected in series with a circuit, never in parallel with it. When an ammeter is connected across a constant-potential source of appreciable voltage, the low internal resistance of the meter bypasses the circuit resistance. This results in the application of the source voltage (or a good portion of it) directly to the meter terminals. The resulting excessive current burns up the meter coil and renders the meter useless until repaired.

Q-11. In what manner are current-measuring instruments connected to a circuit? If you do not know the approximate value of current in the circuit, you should take a reading at the highest range of the ammeter; then you should switch progressively to lower ranges until a suitable reading is obtained. Most ammeter scales indicate the current being measured in increasing values from left to right. If you connect the meter without observing proper polarity, the pointer may be deflected backwards (from right to left). This action often damages the meter movement. You should ensure that the ammeter is always connected so that the current will flow into the negative terminal and out the positive terminal. Figure 3-6 shows various circuit arrangements and the proper ammeter connection methods to measure current in various portions of the circuit.

Q-13. (True or False) The larger the current required to produce full-scale deflection of the meter coil, the better the sensitivity of the meter.

Figure 3-6 Proper ammeter connection

p. 102

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-16 UNCLASSIFIED Good sensitivity is especially important in ammeters to be used in circuits in which small currents flow. As the meter is connected in series with the load, the current flows through the meter. If the internal resistance of the meter is a large portion of the load resistance, an effect known as METER-LOADING will occur. Meter-loading is the condition that exists when the insertion of a meter into a circuit changes the operation of that circuit. This condition is not desirable. The purpose of inserting a meter into a circuit is to allow the measurement of circuit current in the normal operating condition. If the meter changes the circuit operation and changes the amount of current flow, the reading you obtain will be in error. An example of this is shown in figure 3-7.

Q-14. What condition exists when the insertion of a meter into a circuit changes the operation of the circuit? In view A of figure 3-7, the circuit to be tested has an applied voltage of 100 millivolts and a resistance of 100 ohms. The current normally flowing in this circuit is 1 milliampere. In view B, an ammeter that requires 1 milliampere for full-scale deflection and that has an internal resistance of 100 ohms has been inserted. Since 1 milliampere of current flow is shown in view A, you might naturally assume that with the meter inserted into the circuit, a full-scale deflection will occur. You might also assume that the 1 milliampere of circuit current will be measured. However, neither of these assumptions is correct. With the ammeter inserted into the circuit, as shown in view B, the total resistance of the circuit is 200 ohms. With an applied voltage of 100 millivolts, applying Ohm’s law shows the actual current (I circuit) to be 0.5 milliampere.

Figure 3-7 Ammeter loading effect

p. 103

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-17 UNCLASSIFIED I = E R

Icircuit = 100 × 10−3 volts 200 ohms

= 0.5 × 10−3 ampere

= 0.0005 ampere or 0.5 milliampere

Since the meter reads 0.5 milliampere instead of the normal value of current, the meter reveals that a definite loading effect has taken place. In cases such as this, the use of ammeters, which have a lower internal resistance and a better current sensitivity, is desirable.

3.2.4 DC Voltmeter Up to this point, we have been discussing the 100-microampere D’Arsonval movement and its use as an ammeter. However, it can also be used to measure voltage if a MULTIPLIER (high resistance) is placed in series with the moving coil of the meter. For low-voltage instruments, this resistance is physically mounted inside the meter case with the D’Arsonval movement. The series resistance is constructed of a wire-wound resistance that has a low temperature coefficient wound on either a spool or card frame. For high-voltage ranges, the series resistance can be connected externally. A simplified diagram of a voltmeter is shown in figure 3-8.

p. 104

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-18 UNCLASSIFIED

Q-15. What modification is made to the D’Arsonval meter movement to enable the meter to measure voltage? Keep in mind that the D’Arsonval meter movement uses current flow to produce a magnetic field that is proportional to the current. The meter movement is, therefore, an indicator of current flow rather than voltage. The addition of the series resistance is what allows the meter to be calibrated in terms of voltage; that is, the meter movement of a voltmeter operates because of the current flow through the meter, but the scale is marked in volts. For example, the meter movement shown in figure 3-9 has an internal resistance of 100 ohms, requires 100 microamperes for full-scale deflection, and has a voltage drop of 10 millivolts when full-scale deflection is reached. If you were to place this meter directly across a 10-volt source, an excessive current (in milliamperes) would flow. The meter would be destroyed because of the excessive current flowing through the meter movement. This can be seen in the following Ohm’s law application:

Figure 3-8 Internal construction and circuit of a simplified voltmeter

p. 105

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-19 UNCLASSIFIED

Using this equation, you can see that a current through the meter of 100 milliamperes is excessive and will cause damage.

I = E R

= 10 volts 100 ohms

= 100 milliamperes

Since the normal voltage drop for the meter is 10 millivolts at full-scale deflection, some means must be supplied to drop the extra 9.99 volts without applying it directly to the meter. This is done by the addition of a multiplier resistor, as shown in figure 3-9.

Figure 3-9 Use of multiplier resistors with D'Arsonval meter movement

p. 106

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-20 UNCLASSIFIED 3.2.4.1 Extending Voltmeter Ranges The value of series resistance is determined by the current required for full-scale deflection and by the range of the voltages to be measured. Since the current through the meter circuit is directly proportional to the applied voltage, the meter scale can be calibrated directly in volts for a fixed value of series resistance. For example, let’s assume that the basic meter is to be made into a voltmeter with a full-scale deflection of 1 volt. The coil resistance of the basic meter is 100 ohms, and 100 microamperes of current causes full-scale deflection. The resistance (R meter) required to limit the total current in the circuit to 100 microamperes can be found as follows:

Rmeter = E I

1 volt 100 microamperes

= 10 kilohms

Because the meter coil already measures 100 ohms, the series resistance required is equal to 10 kilohms minus 100 ohms, or 9.9 kilohms.

Q-16. What factors determine the value of the multiplier resistor? Multi-range voltmeters use one meter movement. The required resistances are connected in series with the meter by a switching arrangement. A schematic diagram of a multi- range voltmeter with three ranges is shown in figure 3-10. The total meter resistance (R meter) for each of the three ranges, beginning with the 1-volt range, is figured by the application of Ohm’s law, as follows:

p. 107

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-21 UNCLASSIFIED

1-volt range

Rmeter = 1 . 0001 = 10 kilohms

100-volt range

Rmeter = 100 . 0001 = 10 megohm

1,000-volt range

Rmeter = 1000 . 0001 = 10 megohms

The actual value of the multiplying series resistor ( Rseries) for each of these circuits is 100 ohms less than the total resistance. This allows for the resistance of the meter coil (Rcoil).

Figure 3-10 Multi-range voltmeter

p. 108

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-22 UNCLASSIFIED 3.2.4.2 Voltmeter Circuit Connections When voltmeters are used, a primary rule of safety is that such voltage-measuring instruments must always be connected in parallel with (across) a circuit. If you are unsure of the level of the voltage to be measured, take a reading at the highest range of the voltmeter and progressively (step by step) lower the range until a suitable reading is obtained. In many cases, the voltmeter you will be using will not be a center-zero- (0 reading is in the center) indicating instrument. Observing the correct polarity is important when connecting the instrument to the circuit. Voltmeter polarity is the same as for the dc ammeter; that is, current flows from negative to positive.

Q-17. In what manner are voltage-measuring instruments connected to the circuit to be measured? 3.2.4.3 Influence of a Voltmeter in a Circuit The purpose of a voltmeter is to indicate the potential difference between two points in a circuit. When a voltmeter is connected across a circuit, it shunts the circuit. If the voltmeter has a low resistance, it will draw a substantial amount of current. This action lowers the effective resistance of the circuit and changes the voltage reading. When you are making voltage measurements in high-resistance circuits, use a HIGH-RESISTANCE VOLTMETER to prevent the shunting action of the voltmeter. The effect is less noticeable in low-resistance circuits because the shunting effect is less. The problem of voltmeter shunting (sometimes called circuit loading) is illustrated in figure 3-11.

Figure 3-11 Shunting action caused by a voltmeter

p. 109

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-23 UNCLASSIFIED Q-18. When making voltage measurements in a high-resistance circuit, you should always use a voltmeter with what relative value of resistance? In view A of figure 3-11, a source of 150 volts is applied to a series circuit consisting of two 10-kilohm resistors. View A shows the voltage drop across each resistor to be 75 volts. In the 150-volt range, the voltmeter to be used has a total internal resistance of 10 kilohms. View B shows the voltmeter connected across the circuit. The parallel combination of R2 and the meter now present a total resistance of 5 kilohms. Because of the addition of the voltmeter, the voltage drops change to 100 volts across R1 and 50 volts across R2. Notice that this is not the normal voltage drop across R2. Actual circuit conditions have been altered because of the voltmeter.

3.2.4.4 Voltmeter Sensitivity The sensitivity of a voltmeter is given in ohms per volt. It is determined by dividing the sum of the resistance of the meter (R meter), plus the series resistance (Rseries), by the full- scale reading in volts. In equation form, sensitivity is expressed as follows:

sensitivity = Rm + Rs E

This is the same as saying the sensitivity is equal to the reciprocal of the full-scale deflection current. In equation form, this is expressed as follows:

sensitivity = ohms volt

= 1 volt ohms

= 1 ampere

p. 110

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-24 UNCLASSIFIED Therefore, the sensitivity of a 100-microampere movement is the reciprocal of 0.0001 ampere, or 10,000 ohms per volt.

sensitivity = 1 ampere

= 1 . 0001

10,000 ohms per volt

Q-19. What term is used to express the sensitivity of a voltmeter? 3.2.5 Meters Used For Measuring Resistance The two instruments you will use most often to check continuity, or to measure the resistance of a circuit or circuit component, are the OHMMETER and the MEGGER (MEGOHMMETER). The ohmmeter is widely used to measure resistance and to check the continuity of electrical circuits and devices. Its range usually extends to only a few megohms. The megger is widely used for measuring insulation resistance, such as that between a wire and the outer surface of its insulation, and the insulation resistance of cables and insulators. The range of a megger can be extended to more than 1,000 megohms.

Q-20. What instrument is used for measuring the insulation resistance of cables?

p. 111

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-25 UNCLASSIFIED 3.2.5.1 The Ohmmeter A simple ohmmeter circuit is shown in figure 3-12. The ohmmeter consists of the dc milliammeter, discussed earlier in this chapter, and the added features shown below:

• A source of dc potential; and • One or more resistors (one of which is variable).

Q-21. What added features enable a dc milliammeter to function as an ohmmeter?

Figure 3-12 Simple ohmmeter circuit

p. 112

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-26 UNCLASSIFIED The deflection of the pointer of an ohmmeter is controlled by the amount of battery current passing through the moving coil. Before you can measure the resistance of an unknown resistor or electrical circuit, you must calibrate the ohmmeter to be used. If the value of resistance to be measured can be estimated within reasonable limits, select a range on the ohmmeter that will give approximately half-scale deflection when the resistance is inserted between the probes. If you cannot estimate the resistance to be measured, then set the range switch on the highest scale. Whatever range you select, the meter must be calibrated to read zero before the unknown resistance is measured.

To calibrate the meter, you first short the test leads together, as shown in figure 3-12. With the test leads shorted, a complete series circuit exists. The complete series circuit consists of the 3-volt source, the resistance of the meter coil (R meter), the resistance of the zero-adjust rheostat, and the series multiplying resistor (Rseries). The shorted test leads cause current to flow and the meter pointer to deflect.

Notice that the zero point on the ohmmeter scale (as opposed to the zero points for voltage and current) is located at the extreme right side of the scale. With the test leads shorted, the zero-adjust potentiometer is set so that the pointer rests on the zero mark. Therefore, a full-scale deflection indicates zero resistance between the leads.

Q-22. A full-scale deflection on an ohmmeter scale indicates what resistance between the leads? If you change the range on the meter, you must "zero" (calibrate) the meter again to obtain an accurate reading. When you separate the test leads, the pointer of the meter will return to the left side of the scale. This action, as explained earlier, is caused by the restoring force of the spring tension acting on the movable coil assembly. The reading at the left side of the scale indicates an infinite resistance.

After you have adjusted the ohmmeter for zero reading, it is ready to be connected to a circuit to measure resistance. A typical circuit and ohmmeter arrangement is shown in figure 3-13. You must ensure that the power switch of the circuit to be measured is in the de-energized (OFF) position. This prevents the source voltage of the circuit from being applied to the meter, a condition that could cause severe damage to the meter movement.

p. 113

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-27 UNCLASSIFIED

Remember that the ohmmeter is an open circuit when the test leads are separated. To take a resistance reading with a meter, you must provide a path for current flow produced by the meter’s battery. In view A of figure 3-13, the meter is connected at points A and B to produce this path. Connecting these test leads places resistors R1 and R2 in series with the resistance of the meter coil, the zero-adjust potentiometer, and the series multiplying resistor. Since you previously calibrated the meter, the amount of coil movement now depends only on the resistances of R1 and R2.

The addition of R1 and R2 into the meter circuit raises the total series resistance and decreases the current. This decreases the amount of pointer deflection. The pointer comes to rest at a scale reading that indicates the combined resistance of R1 and R2. If you were to replace either R1 or R2, or both, with a resistor having a larger ohmic value, the current flow in the moving coil of the meter would be decreased even more. This would further decrease the pointer deflection, and the scale indication would read a still higher circuit resistance. View B is a simplified version of the circuitry in view A.

From our ohmmeter discussion, two facts should be apparent: (1) Movement of the moving coil is proportional to the amount of current flow, and (2) the scale reading of the ohmmeter is inversely proportional to current flow in the moving coil.

Figure 3-13 Measuring circuit resistance with an ohmmeter.

p. 114

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-28 UNCLASSIFIED The amount of circuit resistance to be measured may vary over a wide range. In some cases, it may only be a few ohms; in other cases, it may be as great as 1 megohm. Scale multiplication features are built into most ohmmeters so that they will indicate any ohmic value being measured and offer the least amount of error. Most ohmmeters are equipped with a selector switch for selecting the multiplication scale desired. For example, view A of figure 3-14 shows a typical meter that has a six-position switch. The positions are marked on the meter in multiples of 10, from R × 1 through R × 100K.

Figure 3-14 Ohmmeter with multiplication switch

p. 115

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-29 UNCLASSIFIED The range used to measure any particular unknown resistance (Rx in view A of figure 3- 14) depends on the approximate ohmic value of the unknown resistance. For instance, the ohmmeter scale of the figure is calibrated in divisions from 0 to infinity. Note that the divisions are easier to read on the right-hand portion of the scale than on the left. For this reason, if Rx is greater than 1,000 ohms and if you are using the R × 1 range, you will be unable to accurately read the indicated resistance. This happens because the combined series resistance of resistors Rx is too large for range R × 1 to allow enough battery current to flow to deflect the pointer away from infinity. You need to turn the range switch to the R × 10 position to obtain the 1,000-ohm reading.

Let’s assume that you have changed the range switch to the R × 10 position and the pointer now deflects to a reading of 375 ohms, as shown in view B of figure 3-14. This would indicate to you that unknown resistance R x has 3,750 (375 times 10) ohms of resistance. The change of range caused the deflection because resistor R × 10 has only 1/10 the resistance of resistor R × 1. Therefore, selecting the smaller series resistance allowed a battery current of sufficient value to cause a readable pointer deflection. If the R × 100 range were used to measure the same 3,750 ohm resistor, the pointer would deflect still further to the 37.5-ohm position, as shown in view C. This increased deflection would occur because resistor R × 100 has only 1/10 the resistance of resistor R × 10.

Q-23. The R × 100 resistance selection on an ohmmeter has what amount of resistance compared to the R × 10 selection? The circuit arrangement in view A of figure 3-14 allows the same amount of current to flow through the moving meter coil. The same amount is allowed to flow whether the meter measures 10,000 ohms on the R × 1 scale, 100,000 ohms on the R × 10 scale, or 1,000,000 ohms on the R × 100 scale.

The same amount of current must always be used to deflect the pointer to a certain position on the scale (midscale position, for example), regardless of the multiplication factor being used. Since the multiplier resistors are of different values, you must always "zero" the meter for each multiplication scale selected. When selecting a range on the ohmmeter, select the one that will result in the pointer coming to rest as close to the midpoint of the scale as possible. This will enable you to read the resistance more accurately because scale readings are more easily interpreted at or near midpoint.

p. 116

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-30 UNCLASSIFIED 3.2.5.2 The Megohmmeter An ordinary ohmmeter cannot be used for measuring multimillion ohm values of resistances, such as those in conductor insulation. To test for such insulation breakdown, you need to use a much higher potential than that supplied by the battery of an ohmmeter. This potential is placed between the conductor and the outside of the insulation. A megger (megohmmeter) is used for these tests. The megger, shown in figure 3-15, is a portable instrument consisting of two main elements: (1) a hand-driven dc generator, which supplies the necessary voltage for making the measurement, and (2) the instrument portion, which indicates the value of the resistance you are measuring. The instrument portion is of the opposed-coil type, as shown in view A. Coils a and b are mounted on movable member c. A fixed angular relationship exists between coils, and they are free to turn as a unit in a magnetic field. Coil b tends to move the pointer counterclockwise, and coil a tends to move it clockwise.

Figure 3-15 Megger internal circuit and external view

p. 117

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-31 UNCLASSIFIED Coil a is connected in series with R3 and unknown resistance Rx. The combination of coil a, R3, and Rx forms a direct series path between the + and − brushes of the dc generator. Coil b is connected in series with R2, and this combination is also connected across the generator. Notice that the movable member (pointer) of the instrument portion of the megger has no restoring springs. Therefore, when the generator is not being operated, the pointer will float freely and may come to rest at any position on the scale.

The guard ring, shown in view A of figure 3-15, shunts any leakage currents to the negative side of the generator. This prevents such current from flowing through coil a and affecting the meter reading.

Q-24. What is the purpose of the guard ring in a megohmmeter? If the test leads are open, no current will flow in coil a. However, current will flow internally through coil b and deflect the pointer to infinity. This reading indicates a resistance too large to measure. When a resistance, such as R x, is connected between the test leads, current also flows in coil a; the pointer tends to move clockwise. At the same time, coil b still tends to move the pointer counterclockwise. Therefore, the moving element, composed of both coils and the pointer, comes to rest at a position in which the two forces are balanced. This position depends upon the value of Rx, which controls the amount of the current in coil a. Because changes in voltage affect both coils in the same proportion, the position of the moving element is independent of the voltage. If you short the test leads together, the pointer will come to rest at zero because the current in coil a is relatively large. Since R3 limits the current, the instrument will not be damaged under these circumstances. The external appearance of one type of megger is shown in view B of figure 3-15.

Most meggers you will use are rated at 500 volts; however, there are other types. Meggers are usually equipped with friction clutches, which are designed to slip if the generator is cranked faster than its rated speed. This prevents the generator speed and output voltage from exceeding rated values. A 1,000-volt generator is available for extended ranges. When an extremely high resistance, such as 10,000 megohms or more, is to be measured, a high voltage is needed to cause enough current flow to actuate the meter movement.

p. 118

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-32 UNCLASSIFIED CAUTION

When using a megger, you can easily be injured or damage equipment if you do not observe the following MINIMUM safety precautions:

• Use meggers on high-resistance measurements only (such as insulation measurements or to check two separate conductors on a cable). • Never touch the test leads while the handle is being cranked. • De-energize and discharge the circuit completely before connecting a megger. • Whenever possible, disconnect the component being checked from other circuitry before using a megger

Q-25. Most meggers you will use are rated at what voltage? Q-26. The development of excessive test voltages is avoided by the use of meggers equipped with what device? 3.3 ELECTRODYNAMOMETER-TYPE METERS The electrodynamometer-type meter differs from the galvanometer types we have just studied in that two fixed coils are used to produce the magnetic field instead of a permanent magnet. Two movable coils are also used in the electrodynamometer meter. The electrodynamometer meter is most commonly found in various types of power meters.

Q-27. What components in an electrodynamometer-type meter movement produce the magnetic field? As shown in figure 3-16, the fixed coils are connected in series and positioned coaxially (in line) with a space between them. The two movable coils are also positioned coaxially and are connected in series. The two pairs of coils (fixed pair and movable pair) are also connected in series with each other. The movable coil is pivot-mounted between the fixed coils. The main shaft on which the movable coils are mounted is restrained by spiral springs that restore the pointer to zero when no current is flowing through the coil. These springs also act as conductors for delivering current to the movable coils. Since these conducting springs are very small, the meter cannot carry a high value of current.

p. 119

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-33 UNCLASSIFIED

Q-28. What is the limiting factor as to the amount of current an electrodynamometer meter movement can handle? 3.3.1 Meter Accuracy The meter is mechanically damped by means of aluminum vanes that move in enclosed air chambers. Although very accurate, electrodynamometer-type meters do not have the sensitivity of the D’Arsonval-type meter movement. For this reason, you will not find them used outside of the laboratory environment to a large extent.

Figure 3-16 Internal construction of an electrodynamometer

p. 120

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-34 UNCLASSIFIED 3.3.2 Meter Movement The primary advantage of the electrodynamometer-type meter movement is that it can be used to measure alternating as well as direct current. If you apply alternating current to the standard galvanometer-type meter, it will not produce a usable reading. Instead, the meter will vibrate at or near the zero reading. On one-half cycle of the ac, the meter is deflected to the left and on the other half cycle to the right. Since the frequencies you will be measuring are 60 hertz or greater, the meter is incapable of mechanically responding at this speed. The result is simply a vibration near the zero point; in addition, no useful reading of voltage or current is obtained. This problem does not exist with the electrodynamometer-type movement. Current flow through the stationary (fixed) coils sets up a magnetic field. Current flow through the moving coils sets up an opposing magnetic field. With two magnetic fields opposing, the pointer deflects to the right. If the current reverses direction, the magnetic fields of both sets of coils will be reversed. With both fields reversed, the coils still oppose each other, and the pointer still deflects to the right. Therefore, no rectifying devices are required to enable the electrodynamometer meter movement to read both ac and dc. Rectifying devices are required for the D’Arsonval-type meter movement to enable it to be used for measuring ac voltages and currents.

Q-29. What is the primary advantage of the electrodynamometer-type meter over the D’Arsonval-type meter?

p. 121

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-35 UNCLASSIFIED 3.3.3 Voltmeter When an electrodynamometer is used as a voltmeter, no problems in construction are encountered because the current required is not more than 0.1 ampere. This amount of current can be handled easily by the spiral springs. When the electrodynamometer is used as a voltmeter, its internal connections and construction are as shown in view A of figure 3-17. Fixed coils a and b are wound of fine wire since the current flow through them will not exceed 0.1 ampere. They are connected directly in series with movable coil c and the series current-limiting resistor.

Figure 3-17 Circuit arrangement of electrodynamometer for use as a voltmeter and an ammeter

p. 122

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-36 UNCLASSIFIED 3.3.4 Ammeter When the electrodynamometer is used as an ammeter, a special type of construction must be used. This is because the large currents that flow through the meter cannot be carried through the moving coils. In the ammeter in view B of figure 3-17, stationary coils a and b are wound of heavier wire to carry up to 5.0 amperes. An inductive shunt (XL) is wired in parallel with the moving coils and permits only a small part of the total current to flow through the moving coil. The current flowing through the moving coil is directly proportional to the total current flowing through the instrument. The shunt has the same ratio of reactance to resistance as the moving coil does. Therefore, the instrument will be reasonably correct at frequencies at which it is used if ac currents are to be measured.

3.3.5 Wattmeter Electric power is measured by means of a wattmeter. This instrument is of the electrodynamometer type. As shown in figure 3-18, it consists of a pair of fixed coils, known as current coils, and a moving coil, called the voltage (potential) coil. The fixed current coils are wound with a few turns of a relatively large conductor. The voltage coil is wound with many turns of fine wire. It is mounted on a shaft that is supported in jeweled bearings so that it can turn inside the stationary coils. The movable coil carries a needle (pointer) that moves over a suitably graduated scale. Coil springs hold the needle at the zero position in the absence of a signal.

Figure 3-18 Simplified electrodynamometer wattmeter circuit

p. 123

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-37 UNCLASSIFIED 3.3.5.1 Wattmeter Connection The current coil of the wattmeter is connected in series with the circuit (load), and the voltage coil is connected across the line. When line current flows through the current coil of a wattmeter, a field is set up around the coil. The strength of this field is in phase with and proportional to the line current. The voltage coil of the wattmeter generally has a high-resistance resistor connected in series with it. The purpose for this connection is to make the voltage-coil circuit of the meter as purely resistive as possible. As a result, current in the voltage circuit is practically in phase with line voltage. Therefore, when voltage is impressed on the voltage circuit, current is proportional to and in phase with the line voltage. Figure 3-19 shows the proper way to connect a wattmeter into a circuit.

3.3.5.2 Wattmeter Errors Electrodynamic wattmeters are subject to errors arising from such factors as temperature and frequency. For example, heat through the coils eventually causes the small springs attached to the pointer to lengthen and lose tension, which produces deflection errors. Large currents through the wattmeter also produce a noticeable deflection error. These errors are caused by the heat (I 2R) loss through coils from the application of high currents. Because of this, the maximum current range of electrodynamic wattmeters is normally restricted to approximately 20 amperes. The voltage range of wattmeters is usually limited to several hundred volts because of heat dissipation within the voltage circuit. However, the voltage range can be extended by the use of voltage multipliers.

Figure 3-19 Wattmeter connection

p. 124

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-38 UNCLASSIFIED Good-quality, portable wattmeters usually have an accuracy of 0.2 to 0.25 percent. You must remember, though, that electrodynamic wattmeter errors increase with frequency. For the higher frequency and power ranges, special types of wattmeters are made specifically for those ranges. We will discuss two such wattmeters in chapter 5 of this module

3.3.5.3 Wattmeter Overloads The wattmeter consists of two circuits, either of which will be damaged if too much current passes through them. You should be especially aware of this fact because the reading on the instrument will not tell you whether or not the coils are being overheated. If an ammeter or voltmeter is overloaded, the pointer will indicate beyond the upper limit of its scale. In the wattmeter, both the current and potential circuit may carry such an overload that their insulations burn; yet the pointer may be only part of the way up the scale. This is because the position of the pointer depends upon the power factor of the circuit as well as upon the voltage and current. Therefore, a low power-factor circuit will provide a very low reading on the wattmeter. The reading will be low, even when the current and voltage circuits are loaded to the maximum safe limit. The safe rating for each wattmeter is always distinctly rated, not in watts, but in volts and amperes.

3.4 TECHNIQUES FOR METER USE We have considered the more common meters; now let’s consider some of the techniques employed in their use. The techniques suggested here are not all-inclusive. You will find, as you develop your technical skills, other variations and techniques in use. Consider the techniques for measuring current in a circuit. You can accomplish this by placing an ammeter in series with the circuit or by measuring the voltage across a resistor of known value and using Ohm’s law to figure current. This last technique has the advantage of eliminating the necessity of opening the circuit to connect the ammeter.

3.4.1 Continuity Tests Open circuits are those in which the flow of current is interrupted by a broken wire, defective switch, or any means by which the current cannot flow. The test used to detect open circuits (or to see if the circuit is complete or continuous) is continuity testing.

An ohmmeter (which contains its own batteries) is excellent for use in a continuity test. Normally, continuity tests are performed in circuits where the resistance is very low, such as the resistance of a copper conductor. An open is indicated in these circuits by a very high or infinite resistance between two continuously connected points.

p. 125

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-39 UNCLASSIFIED Figure 3-20 shows a continuity test of a cable that connects two electronic units. Notice that both plugs are disconnected and the ohmmeter is in series with conductor D under test. The power should be off. When checking conductors A, B, and C (connection of ohmmeter to conductors not shown), the current from the ohmmeter flows through plug 2 (female) through conductor A, B, or C to plug 1 (female). From plug 1, current passes through the jumper to the chassis, which is "grounded" to the ship’s structure. The metal structure serves as the return path to the chassis of unit 2 and completes the circuit through the series-connected ohmmeter. The ohmmeter indicates a low resistance because no break exists in conductors A, B, or C. However, checking conductor D reveals an open. The ohmmeter is shown indicating maximum resistance because current cannot flow in an open circuit. With an open circuit, the ohmmeter needle is all the way to the left since it is a series-type ohmmeter (reads right to left).

Where conditions are such that the ship’s structure cannot be used as the return path, one of the other conductors (known to be good) may be used. For example, to check D, you can connect a jumper from pin D to pin A of plug 1 (female) and the ohmmeter leads to pins D and A of plug 2 (female). This technique will also reveal the open in the circuit.

Figure 3-20 Continuity test

p. 126

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-40 UNCLASSIFIED 3.4.2 Testing For Grounds Grounded circuits are caused by some conducting part of the circuit making contact either directly or indirectly with the metallic structure of the ship. Grounds can have many causes. The two most common are the fraying of insulation from a wire and moisture-soaked insulation. The fraying of insulation from a wire allows bare wire to come into contact with the metal ground. Moisture-soaked insulation causes reduced insulation resistance (also classified as a ground).

Grounds are usually indicated by blown fuses or tripped circuit breakers. Blown fuses or tripped circuit breakers, however, can also result from a short circuit other than a ground. A high-resistance ground can also occur when current is increased significantly but not enough to rupture the fuse or trip the circuit breaker.

CAUTION

Before testing any circuit, ensure the circuit under test has been de-energized and checked with a safety shorting probe.

In testing for grounds, you may use a megger or an ohmmeter. Measuring the resistance to ground from points in a circuit determines if the point is grounded. Referring again to figure 3-20, you can see one possible means of testing a cable for grounds. If the jumper is removed from pin D of plug 1 (female), a test for ground can be made for each conductor in the cable. You can do this by connecting one meter lead to ground and the other to each of the pins of either of the plugs. A low resistance indicates that some part of that conductor or one of the plug assemblies is grounded. Both plugs must be removed from their units; if only one plug is removed, a false indication is possible because a conductor may be grounded through the unit.

3.4.3 Testing For Shorts A short circuit, other than a grounded one, is one where two conductors touch each other directly or through another conducting element. Two conductors with frayed insulation may touch and cause a short. Too much solder on the pin of a connector may short to the adjacent pin. In a short circuit, enough current may or may not flow to blow a fuse or open a circuit breaker. A short may occur between two cables carrying signals but might not be indicated by a blown fuse.

Shorts occur in many components, such as transformers, motor windings, and capacitors. The major test method used to detect shorts in such components is to measure resistance. The indicated resistance is then compared with the resistance given on schematics or in the equipment technical manuals to determine whether the measured value is within specifications.

p. 127

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-41 UNCLASSIFIED An ohmmeter is the device used to check for shorts. You can use the ohmmeter to detect a short between two conductors by measuring the resistance between them (be sure electrical power has been disconnected). A low resistance reading indicates a short. You can test the circuit in figure 3-20 for a short by first removing the jumper and disconnecting both plugs; you then measure the resistance between the two suspended conductors.

WARNING

The following section discusses voltage measurements on live circuits. BE SURE YOU ALWAYS FOLLOW PRESCRIBED SAFETY RULES WHEN MEASURING VOLTAGES.

3.4.4 Voltage Tests Voltage tests must be made with the power applied; therefore, the prescribed safety precautions must be followed to prevent injury to personnel and damage to the equipment. You will find in your maintenance work that the voltage test is of utmost importance. It is used not only in isolating casualties to major components but also in the maintenance of subassemblies, units, and circuits. Before checking a circuit voltage, you should check the voltage of the power source to be sure that the normal voltage is being applied to the circuit.

The voltmeter is used for voltage tests. In using the voltmeter, make certain that the meter used is designed for the type of current (ac or dc) to be tested and has a scale with a suitable range. Since defective parts in a circuit can cause higher than normal voltages to be present at the point of test, the highest voltmeter range available should be used first. Once you have obtained a reading, determine if a lower scale can be used that will cause no damage to the meter movement. If so, use the lower scale. This provides a more accurate reading.

Another consideration in the circuit voltage test is the resistance and current in the circuit. A low resistance in a high-current circuit could result in considerable voltage drop, whereas the same resistance in a low-current circuit may be minimal. Abnormal resistance in part of a circuit can be checked with either an ohmmeter or a voltmeter. Where practical, an ohmmeter should be used because the test is then carried out with a "dead" circuit.

The majority of the electronic circuits you will encounter in equipment will be low- current circuits, and most voltage readings will be direct current. Also, many of the schematics will indicate the voltages at various test points. Therefore, if you suspect that a certain stage is defective, you can check the voltage by connecting a voltmeter from the test point to ground. If the suspected stage is not defective, the voltmeter readings should match the voltages given on the schematic.

p. 128

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-42 UNCLASSIFIED Some technical manuals also contain voltage charts on which all the voltage measurements are tabulated. These charts usually indicate the sensitivity of the meter (for example, 20,000 ohms/volt) used to obtain the voltage readings for the chart. To obtain comparable results, you must use a voltmeter of the same sensitivity (or greater) as that specified. Make certain that the voltmeter is not "loading" the circuit while taking a measurement. If the meter resistance is not considerably higher than the circuit resistance, the reading will be markedly lower than the true circuit voltage because of the voltmeter’s loading effect. (To calculate meter resistance, multiply the rated ohms-per- volt sensitivity value of the meter by the scale in use. For example, a 1,000-ohms-per-volt meter set to the 300-volt scale will have a resistance of 300,000 ohms.)

3.4.5 Resistance Tests Before checking the resistance of a circuit or of a part, make certain that the power has been turned off. Also make sure capacitors in the associated circuit are fully discharged. To check continuity, always use the lowest ohmmeter range. If the highest range is used, the meter may indicate zero, even though appreciable resistance is present in the circuit. Conversely, to check a high resistance, use the highest scale since the lower range scale may indicate infinity, even though the resistance is less than a megohm. In making resistance tests, you must remember that even though the external ohmmeter leads are connected in parallel with the circuit to be measured, the internal meter circuitry is electrically connected in series.

In making resistance tests, take into account that other circuits containing resistances and capacitances may be in parallel with the circuit to be measured. Erroneous conclusions may be drawn from readings obtained in such cases. Remember, a capacitor blocks the dc flow from the ohmmeter. To obtain an accurate reading when other parts are connected across the suspected circuit, disconnect one end of the circuit to be measured from the equipment. For example, many of the resistors in major components and subassemblies are connected across transformer windings. To obtain a valid resistance measurement, you must isolate the resistors to be measured from the shunt resistances of the coils of the transformers.

Resistance tests are also used to check a component for grounds. In these tests, the component to be tested should be disconnected from the rest of the circuit so that no normal circuit ground will exist. Dismounting the component to be checked is not necessary. The ohmmeter is set for a high-resistance range. Then the ohmmeter is connected between ground and each electrically separate circuit of the component being tested. Any resistance reading less than infinity indicates at least a partial ground. You can also check capacitors suspected of being short-circuited by measuring the resistance. To check a capacitor suspected of being open, temporarily shunt a known good capacitor then recheck the performance of the circuit.

p. 129

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-43 UNCLASSIFIED CAUTION

To avoid possible damage to equipment during resistance tests, observe the following precautions:

• Always connect an ammeter in series—never in parallel. • Connect a voltmeter in parallel. • Never connect an ohmmeter to a live circuit. • Observe polarity when using a dc ammeter or a dc voltmeter. • View meters directly from the front. When viewed from an angle off to the side, an incorrect reading will result because of OPTICAL PARALLAX. (Parallax was covered in NEETS, Module 3, Introduction to Circuit Protection, Control, and Measurement.) • Always choose an instrument suitable for the measurement desired. • Select the highest range first and then switch to the proper range. • In using a meter, choose a scale that will result in an indication as near midscale as possible. • Do not mount or use instruments in the presence of a strong magnetic field. • Remember, a low internal resistance voltmeter (low sensitivity) may shunt the circuit being measured and result in incorrect readings.

p. 130

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-44 UNCLASSIFIED 3.5 SUMMARY The important points of this chapter are summarized in the following paragraphs. You should be familiar with these points before continuing with your study of test equipment.

A permanent-magnet, moving-coil meter movement (D’ARSONVAL movement) uses the interaction of magnetic fields to produce movement.

DAMPING is used to smooth out the vibration and to help prevent overshooting of the meter pointer.

ELECTRODYNAMOMETER movements are usually used in wattmeters. They operate much like the D’Arsonval meter movement, except field coils are used instead of a permanent magnet. Electrodynamometer movements measure either ac or dc without the use of a rectifier.

A SHUNT is a physically large, low-resistance conductor connected in parallel with the meter terminals. It carries the majority of the load current so that only a small portion of the total current will flow through the meter coil.

An AMMETER measures current and is always connected in series with the circuit being measured. An ammeter should have a low resistance so that the effect of the ammeter on the circuit will be kept to a minimum.

VOLTMETERS are used to measure voltage and are always connected in parallel with the circuit being measured. A voltmeter should have a high resistance compared to the circuit being measured to minimize the loading effect. Voltmeter sensitivity is expressed in ohms per volt.

OHMMETERS are used to measure resistance and to check continuity. An ohmmeter is electrically connected in series with the resistance being measured. The ohmmeter range, which allows a midscale deflection, should be used.

A MEGOHMMETER (MEGGER) is used to measure very high resistance, such as the insulation of wiring.

A WATTMETER is usually an electrodynamometer and is used to measure power.

A CONTINUITY TEST is accomplished with an ohmmeter. This test is used to check for opens (or to see if the circuit is complete or continuous).

GROUNDED CIRCUITS are caused by some conducting part of the circuit making contact either directly or indirectly with the metallic structure of the ship or chassis. In testing for grounds, you may use either an ohmmeter or a megger.

p. 131

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-45 UNCLASSIFIED A SHORT CIRCUIT, other than a grounded one, is where two conductors touch each other directly or through another conducting element. An ohmmeter is used to test for shorts.

p. 132

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-46 UNCLASSIFIED ANSWERS TO QUESTIONS Q1. THROUGH Q29.

A-1. Self-excited.

A-2. Phosphor bronze ribbons.

A-3. The pointer arrangement and the light and mirror arrangement.

A-4. Coil balance.

A-5. Hairspring.

A-6. Hairspring.

A-7. Makes it possible to have a more linear scale than if the poles were flat.

A-8. Shunt.

A-9. Zero-temperature coefficient.

A-10. Midscale.

A-11. In series.

A-12. Negative, positive.

A-13. False.

A-14. Meter-loading.

A-15. A multimeter (high resistance) is placed in series with the coil of the meter.

A-16. The current required for full-scale deflection, and the range of the voltage to be measured.

A-17. In parallel.

A-18. High.

A-19. Ohms per volt.

A-20. Megohmmeter (megger).

p. 133

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

3-47 UNCLASSIFIED A-21. 1. A source of dc potential. 2. One or more resistors (one of which is variable).

A-22. Zero.

A-23. 1/10.

A-24. Shunts leakage current, which prevents false readings.

A-25. 500.

A-26. Friction clutches.

A-27. Fixed coils.

A-28. Size of spiral conducting.

A-29. The electrodynamometer-type meter can be used to measure both ac and dc currents.

Chapter 6 The Oscilloscope and Spectrum Analyzer

p. 226

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-1 UNCLASSIFIED 6 THE OSCILLOSCOPE AND SPECTRUM ANALYZER LEARNING OBJECTIVES

Upon completing this chapter, you should be able to:

1. Describe the purpose of the CRT used in the oscilloscope. 2. Explain the operation of an oscilloscope. 3. Describe the purpose of the controls and indicators found on an oscilloscope. 4. Describe the proper procedure for using a dual-trace oscilloscope. 5. Describe the accessory probes available for use with a dual-trace oscilloscope. 6. Explain the operation of the spectrum analyzer. 7. Describe the purpose of the controls and indicators found on the spectrum analyzer.

6.1 INTRODUCTION One of the most widely used pieces of electronic test equipment is the OSCILLOSCOPE. An oscilloscope is used to show the shape of a video pulse appearing at a selected equipment test point. Although some oscilloscopes are better than others in accurately showing video pulses, all function in fundamentally the same way. If you learn how one oscilloscope operates, you will be able to learn others.

As you will learn in this chapter, there are many different types of oscilloscopes - varying in complexity from the simple to the complex. Before we get into our discussion of the dual-trace oscilloscope, we will first present a general overview of basic single-trace oscilloscope operation. Shortly, we will see how oscilloscopes use a CATHODE-RAY TUBE (CRT) in which controlled electron beams are used to present a visible pattern of graphical data on a fluorescent screen.

Another piece of test equipment used is the SPECTRUM ANALYZER. This test equipment is used to sweep over a band of frequencies to determine what frequencies are being produced by a specific circuit under test, and then the amplitude of each frequency component. An accurate interpretation of the display will allow you to determine the efficiency of the equipment being tested.

p. 227

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-2 UNCLASSIFIED 6.2 CATHODE-RAY TUBES A detailed discussion of CATHODE-RAY TUBES (CRTs) is presented in NEETS, Module 6, Electronic Emission, Tubes, and Power Supplies. Before continuing with your study of CRTs in this section, you may want to review chapter 2 of that module.

Cathode-ray tubes used in oscilloscopes consist of an ELECTRON GUN, a DEFLECTION SYSTEM, and a FLUORESCENT SCREEN. All of these elements are enclosed in the evacuated space inside the glass CRT. The electron gun generates electrons and focuses them into a narrow beam. The deflection system moves the beam horizontally and vertically across the screen. The screen is coated with a phosphorous material that glows when struck by the electrons. Figure 6-1 shows the construction of a CRT.

6.2.1 Electron Gun The ELECTRON GUN consists of a HEATER and a CATHODE to generate electrons, a CONTROL GRID to control brightness by controlling electron flow, and two ANODES (FIRST and SECOND). The main purpose of the first (FOCUSING) anode is to focus the electrons into a narrow beam on the screen. The second (ACCELERATING) anode accelerates the electrons as they pass. The control grid is cylindrical and has a small opening in a baffle at one end. The anodes consist of two cylinders that contain baffles (or plates) with small holes in their centers.

Figure 6-1 Construction of a CRT

p. 228

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-3 UNCLASSIFIED Q-1. What element controls the number of electrons striking the screen? Q-2. What element is controlled to focus the beam? 6.2.1.1 Cathode and Control Grid As in most conventional electron tubes, the cathode is indirectly heated and emits a cloud of electrons. The control grid is a hollow metal tube placed over the cathode. A small opening is located in the center of a baffle at the end opposite the cathode. The control grid is maintained at a negative potential with respect to the cathode to keep the electrons bunched together.

A high positive potential on the anodes pulls electrons through the hole in the grid. Because the grid is near the cathode, it can control the number of electrons that are emitted. As in an ordinary electron tube, the negative voltage of the grid can be varied either to control electron flow or stop it completely. The brightness (intensity) of the image on the fluorescent screen is determined by the number of electrons striking the screen. This is controlled by the voltage on the control grid.

6.2.1.2 Electrostatic Lenses and Focusing The electron beam is focused by two ELECTROSTATIC FIELDS that exist between the control grid and first anode and between the first and second anodes.

Figure 6-2 shows you how electrons move through the electron gun. The electrostatic field areas are often referred to as LENSES because the fields bend electron streams in the same manner that optical lenses bend light rays. The first electrostatic lens cause the electrons to cross at the first focal point within the field. The second lens bend the spreading streams and return them to a new, second focal point at the CRT.

Q-3. Why are the electrostatic fields between the electron gun elements called lenses?

p. 229

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-4 UNCLASSIFIED

Figure 6-2 also shows the relative voltage relationships on the electron-gun elements. The cathode (K) is at a fixed positive voltage with respect to ground. The grid is at a variable negative voltage with respect to the cathode. A fixed positive voltage of several thousand volts is connected to the second (accelerating) anode. The potential of the first (focusing) anode is less positive than the potential of the second anode. The first anode can be varied to place the focal point of the electron beam on the screen of the tube. Control-grid potential is established at the proper level to allow the correct number of electrons through the gun for the desired image intensity.

Q-4. What is the function of the second anode?

Figure 6-2 Formation of an electron beam

p. 230

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-5 UNCLASSIFIED 6.2.2 Electron Beam-Deflection System The electron beam is developed, focused, and accelerated by the electron gun. The beam appears on the screen of the CRT as a small, bright dot. If the beam is left in one position, the electrons will soon burn away the illuminating coating in that one area. To be of any use, the beam must be able to move. As you have studied, an electrostatic field can bend the path of a moving electron.

As you have seen in the previous illustrations, the beam of electrons passes through an electrostatic field between two plates. You should remember that electrons are negatively charged and that they will be deflected in the direction of the electric force (from negative to positive). This deflection causes the electrons to follow a curved path while in the electrostatic field.

When the electrons leave the electrostatic field, they will take a straight path to the screen at the angle at which they left the field. Because they were all deflected equally, the electrons will be traveling toward the same spot. Of course, the proper voltages must exist on the anodes to produce the electrostatic field. Changing these voltages changes the focal point of the beam and causes the electron beam to strike the CRT at a different point.

6.2.2.1 Factors Influencing Deflection The ANGLE OF DEFLECTION (the angle the outgoing electron beam makes with the CRT center line axis between the plates) depends on the following factors:

• Length of the deflection field; • Spacing between the deflection plates; • The difference of potential between the plates; and • The accelerating voltage on the second anode.

LENGTH OF DEFLECTION FIELD - As shown in figure 6-3, a long field (long deflection plates) has more time to exert its deflecting forces on an electron beam than does a shorter field (short deflection plates). Therefore, the longer deflection plates can bend the beam to a greater deflection angle.

p. 231

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-6 UNCLASSIFIED

Q-5. What effect do longer deflection plates have on the electron beam? SPACING BETWEEN PLATES - As shown in figure 6-4, the closer together the plates, the more effect the electric force has on the deflection angle of the electron beam.

Figure 6-3 Factors influencing length of field Figure 6-4 Spacing between plates

p. 232

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-7 UNCLASSIFIED DIFFERENCE OF POTENTIAL - The potential on the plates (figure 6-5) can be varied to cause a wider or narrower deflection angle. The greater the potential, the wider the deflection angle.

Q-7. Is the deflection angle greater with higher or lower potential on the plates? BEAM ACCELERATION - The faster the electrons are moving, the smaller their deflection angle will be, as shown in figure 6-6.

Q-8. Is the deflection angle greater when the beam is moving faster or slower?

Figure 6-5 Differences of potential Figure 6-6 Beam acceleration

p. 233

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-8 UNCLASSIFIED 6.2.2.2 Vertical and Horizontal Plates If two sets of deflection plates are placed at right angles to each other inside a CRT (figure 6-7), the electron beam can be controlled in any direction. By varying the potential of the vertical-deflection plates, you can make the spot (beam) on the face of the tube move vertically. The distance the beam moves will be proportional to the change in potential difference between the plates. Changing the potential difference between the horizontal-deflection plates will cause the beam to move a given distance from one side to the other. Directions other than up-down and left-right are achieved by a combination of horizontal and vertical movement.

As shown in figure 6-8, position X of the beam is in the center. It can be moved to position Y by going up 2 units and then right 2 units. Movement of the beam is the result of the simultaneous action of both sets of deflection plates. The electrostatic field between the vertical plates moves the electrons up an amount proportional to 2 units on the screen. As the beam passes between the horizontal plates, it moves to the right an amount proportional to 2 units on the screen.

Figure 6-7 Deflection plate arrangement

p. 234

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-9 UNCLASSIFIED

If the amount of deflection from the left and down occurred so that each set of plates acted at the same time, the picture would be like the one in view A of figure 6-9. For example, if the vertical plates moved the beam downward (starting from point X) at the rate of 3 units per second and the horizontal plates moved it to the left at the rate of 1 unit per second, both movements would have been completed in 1 second at point Y. The result would be a straight line.

Figure 6-8 Beam movement on the CRT Figure 6-9 Deflection of the beam

p. 235

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-10 UNCLASSIFIED In view B, the potentials on the vertical and horizontal plates change at the same rate. In the same time period, say 1 second, both plates move the beam 1 unit. The horizontal plates have completed their task at the end of 1 second, but the vertical plates have moved the beam only one-third of the required distance. In this case, the picture in view B would appear on the screen.

6.2.2.3 Beam-deflection Plate Action Recall from your study of chapter 2 of this module that waveforms are described in terms of amplitude versus time. You have just seen how the movement of the CRT beam depends on both potential (amplitude) and time.

Q-9. Waveforms are described in terms of what two functions? VERTICAL-DEFLECTION PLATES - We will use figure 6-10 to explain the action of the vertical-deflection plates in signal amplitude measurements. As this discussion begins, remember that vertical-deflection plates are used to show amplitude of a signal, and horizontal-deflection plates are used to show time and/or frequency relationships.

Figure 6-10 Amplitude versus time

p. 236

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-11 UNCLASSIFIED 1. From T0 to T1, the vertical plates maintain their static difference in potential and the beam stays at 0 units; the T0 to T1 change causes an increasing potential difference in the horizontal plates, and the beam moves 1 unit to the right.

2. At T1, a positive potential difference change in the vertical plates occurs, which causes the beam to move up (instantaneously) 2 units. This vertical (amplitude) beam location is maintained from T1 to T4; horizontal beam movement continues moving to the right as 3 units of time pass.

3. At T4, an instantaneous negative change in potential of 4 units in amplitude occurs, and the beam moves from +2 to -2 units.

4. From T4 to T7, the beam remains at -2 units. During this time period, the beam continues moving horizontally to the right, indicating the passage of time.

5. At T7, a positive increase of amplitude occurs, and the beam moves vertically from -2 to 0 units. From T7 to T8, no change occurs in vertical beam movement; however, horizontal movement continues with time.

The vertical-plate potential difference follows the voltage of the waveform. The horizontal-plate potential follows the passage of time. Together, they produce the image (trace) produced on the screen by the moving beam.

Q-10. The vertical-deflection plates are used to reproduce what function? Q-11. The horizontal-deflection plates are used to produce what function? HORIZONTAL-DEFLECTION PLATES - Now let's look at horizontal-deflection action. Assume that the resistance of the potentiometer shown in figure 6-11 is spread evenly along its length. When the arm of the potentiometer is at the middle position, the same potential exists on each plate. Since there is zero potential difference between the plates, an electrostatic field is not moved downward at a uniform rate; the right plate will become more positive than the left (you are looking down through the top of the CRT). The electron beam will move to the right from screen point 0 through points 1, 2, 3, and 4 in equal time intervals.

p. 237

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-12 UNCLASSIFIED

If the potentiometer arm is moved at the same rate in the opposite direction, the right plate will decrease in positive potential until the beam returns to the 0 position. At that point, the potential difference between the plates is again zero. Moving the arm toward the other end of the resistance causes the left plate to become more positive than the right, and the beam moves from screen points 0 through 4. If the movement of the potentiometer arm is at a uniform (linear) rate, the beam moves at a uniform rate.

Notice that the ends of the deflection plates are bent outward to permit wide-angle deflection of the beam. The vertical plates are bent up and down in the same manner.

Q-12. Why are the ends of the deflection plates bent outward? For ease of explanation, the manual movement of the potentiometer arm is satisfactory to introduce you to horizontal beam movement. However, in the oscilloscope this is not how horizontal deflection is accomplished. Beam movement voltages are produced much faster by sawtooth circuitry. You may want to review the sawtooth generation section in NEETS, Module 9, Introduction to Wave-Generation and Wave-Shaping Circuitry before continuing. Nearly all oscilloscopes with electrostatic deflection apply a sawtooth voltage to the horizontal plates to produce horizontal deflection of the beam, as shown in figure 6-12.

Figure 6-11 Horizontal plates (top view)

p. 238

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-13 UNCLASSIFIED

In the figure, the sawtooth generator replaces the potentiometer and is connected to both horizontal plates of the CRT. At the reference line, the potential on both plates is equal. Below the line, the left plate is more positive and the right plate is less positive. This causes the beam to move left. Above the line, the right plate is made more positive than the left and the beam moves to the right. The waveform amplitude causes a uniform movement of the beam across the screen (called TRACE). RETRACE time, shown at the trailing edge of the waveform, quickly deflects the beam back to the starting point.

6.2.3 CRT Graticule A GRATICULE was used in our previous discussion in figure 6-10. It is simply a calibrated scale (made of clear plastic) of amplitude versus time that is placed on the face of the CRT.

The graticule can be used to determine the voltage of waveforms because the DEFLECTION SENSITIVITY of a CRT is uniform throughout the vertical plane of the screen. Deflection sensitivity states the number of inches, centimeters, or millimeters a beam will be deflected for each volt of potential difference applied to the deflection plates. It is directly proportional to the physical length of the deflection plates and their distance from the screen and inversely proportional to the distance between the plates and to the second-anode voltage. Deflection sensitivity is a constant that is dependent on the construction of the tube.

Deflection sensitivity for a given CRT might typically be 0.2 millimeters per volt. This means the spot on the screen will be deflected 0.2 millimeters (about 0.008 inch) when a difference of 1 volt exists between the plates. Sometimes the reciprocal of deflection sensitivity (called DEFLECTION FACTOR) is given. The deflection factor for the example given would be 125 volts per inch (1/0.008).

Figure 6-12 Sawtooth generator

p. 239

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-14 UNCLASSIFIED Q-13. What term is used to describe the reciprocal of deflection sensitivity of a scope? In the above example, 125 volts applied between one set of plates would deflect the beam 1 inch on the screen. This means that the deflection caused by small signals would likely not be observed. For this reason, the deflection plates are connected to amplifiers that magnify the signals applied to the vertical input of the scope.

Assume, for example, that a peak-to-peak value of a known voltage applied to the oscilloscope indicates that each inch marking on the graticule is equal to 60 volts. Each of the 10 subdivisions will, therefore, equal a value of 6 volts. Most oscilloscopes have ATTENUATOR controls to decrease or GAIN controls to increase the strength of a signal before it is placed on the deflection plates. Attenuator and gain settings must not be disturbed after the calibration has been made. For maximum accuracy, you should recalibrate the graticule each time a voltage is to be measured.

6.2.4 CRT Designations Cathode-ray tubes are identified by a tube number, such as 2AP1, 2BP4, or 5AP1A. The first number identifies the diameter of the tube face. Typical diameters are 2 inches, 5 inches, and 7 inches. The first letter designates the order in which a tube of a given diameter was registered. The letter-digit combination indicates the type of phosphor (glowing material) used on the inside of the screen. Phosphor P1, which is used in most oscilloscopes, produces a green light at medium PERSISTENCE. Persistence refers to the length of time the phosphor glows after the electron beam is removed. P4 provides a white light and has a short persistence. If a letter appears at the end, it signifies the number of the modification after the original design.

p. 240

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-15 UNCLASSIFIED 6.3 OSCILLOSCOPE CONTROL COMPONENTS Although the CRT is a highly versatile device, it cannot operate without control circuits. The type of control circuits required depends on the purpose of the equipment in which the CRT is used.

There are many different types of oscilloscopes. They vary from relatively simple test instruments to highly accurate laboratory models. Although oscilloscopes have different types of circuits, most can be divided into the basic sections shown in figure 6-13: (1) a CRT, (2) a group of control circuits that control the waveform fed to the CRT, (3) a power supply, (4) sweep circuitry, and (5) deflection circuitry.

Q-14. List the circuits that all oscilloscopes have in common.

Figure 6-13 Block diagram of an oscilloscope

p. 241

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-16 UNCLASSIFIED Figure 6-14 is a drawing of the front panel of a dual-trace, general-purpose oscilloscope. Oscilloscopes vary greatly in the number of controls and connectors. Usually, the more controls and connectors, the more versatile the instrument. Regardless of the number, all oscilloscopes have similar controls and connectors. Once you learn the fundamental operation of these common controls, you can move with relative ease from one model of oscilloscope to another. Occasionally, controls that serve similar functions will be labeled differently from one model to another. However, you will find that most controls are logically grouped and that their names usually indicate their function.

The oscilloscope in figure 6-14 is called DUAL-TRACE because it can accept and display two vertical signal inputs at the same time - usually for comparison of the two signals or one signal and a reference signal. This scope can also accept just one input. In this case, it is used as a SINGLE-TRACE OSCILLOSCOPE. For the following discussion, we will consider this to be a single-trace oscilloscope. The oscilloscope in the figure is commonly used in the fleet. You are likely to use this one (model AN/USM- 425) or one very similar to it. Let's now look at the front panel controls.

Figure 6-14 Dual-trace oscilloscope

p. 242

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-17 UNCLASSIFIED 6.3.1 Components Used to Display the Waveform The CRT DISPLAY SCREEN is used to display the signal (figure 6-15). It allows you to make accurate measurements using the vertical and horizontal graticules, as discussed earlier.

Figure 6-15 CRT display and graticule

p. 243

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-18 UNCLASSIFIED 6.3.2 Components Used to Adjust CRT Display Quality The controls in figure 6-16 allow you to adjust for a clear signal display. They also allow you to adjust the display position and magnify the horizontal trace by a factor of 10 (X10). Keep in mind that the controls may be labeled differently from one model to another, depending on the manufacturer. Refer to figure 6-16 as you study the control descriptions in the next paragraphs.

6.3.2.1 INTEN (Intensity) Control The INTEN (intensity) control (sometimes called BRIGHTNESS) adjusts the brightness of the beam on the CRT. The control is rotated in a clockwise direction to increase the intensity of the beam and should be adjusted to a minimum brightness level that is comfortable for viewing.

Figure 6-16 Quality adjustment for CRT display

p. 244

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-19 UNCLASSIFIED 6.3.2.2 FOCUS and ASTIG (Astigmatism) Controls The FOCUS control adjusts the beam size. The ASTIG (astigmatism) control adjusts the beam shape. The FOCUS and ASTIG controls are adjusted together to produce a small, clearly defined circular dot. When displaying a line trace, you will use these same controls to produce a well-defined line. Figure 6-17, view A, shows an out-of-focus beam dot. View B shows the beam in focus. Views C and D show out-of-focus and in-focus traces, respectively.

6.3.2.3 TRACE ROTATION Control The TRACE ROTATION control (figure 6-16) allows for minor adjustments of the horizontal portion of the trace so that you can align it with the horizontal lines on the graticule.

Figure 6-17 Effects of FOCUS and ASTIG (astigmatism) controls

p. 245

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-20 UNCLASSIFIED 6.3.2.4 BEAM FINDER Control Occasionally, the trace will actually be located off the CRT (up or down or to the left or right) because of the orientation of the deflection plates. When pushed, the BEAM FINDER (figure 6-16) pulls the beam onto the screen so that you can use the horizontal and vertical POSITION controls to center the spot.

6.3.2.5 Horizontal and Vertical POSITION Controls The horizontal and vertical POSITION controls (figure 6-16) are used to position the trace. Because the graticule is often drawn to represent a graph, some oscilloscopes have the positioning controls labeled to correspond to the X and Y axes of the graph. The X axis represents horizontal movement; the Y axis represents the vertical movement. Figure 6-18 shows the effects of positioning controls on the trace.

In view A, the horizontal control has been adjusted to move the trace too far to the right; in view B, the trace has been moved too far to the left. In view C, the vertical POSITION control (discussed later) has been adjusted to move the trace too close to the top; in view D, the trace has been moved too close to the bottom. View E (figure 6-18) shows the trace properly positioned.

Figure 6-18 Effects of horizontal and vertical controls

p. 246

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-21 UNCLASSIFIED 6.3.2.6 10X MAG (Magnifier) Switch The 10X MAG (magnifier) switch (figure 6-16) allows you to magnify the displayed signal by a factor of 10 in the horizontal direction. This ability is important when you need to expand the signal to evaluate it carefully.

6.3.3 Components Used to Determine the Amplitude of a Signal We will now discuss the dual-trace components of the scope. You will use these components to determine the amplitude of a signal. Notice in figure 6-19 that the highlighted section at the upper left of the scope looks just the same as the section at the lower left of the scope. This reveals the dual-trace capability section of the scope. The upper left section is the CH (channel) 1 input and is the same as the CH 2 input at the lower left. An input to both inputs at the same time will produce two independent traces on the CRT and use the dual-trace capability of the scope.

Figure 6-19 Components that determine amplitude

p. 247

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-22 UNCLASSIFIED For purposes of this introductory discussion, we will present only CH (channel) 1. You should realize that the information presented also applies to CH 2.

6.3.3.1 Vertical POSITION Control The vertical POSITION control allows you to move the beam position up or down, as discussed earlier.

6.3.3.2 Input Connector The vertical input (or signal input) jack connects the signal to be examined to the vertical-deflection amplifier. Some oscilloscopes may have two input jacks, one labeled AC and the other labeled DC. Other models may have a single input jack with an associated switch, such as the AC GRD DC switch in figure 6-19. This switch is used to select the ac or dc connection. In the DC position, the signal is connected directly to the vertical-deflection amplifier; in the AC position, the signal is first fed through a capacitor. Figure 6-20 shows the schematic of one arrangement.

Figure 6-20 Vertical input arrangement

p. 248

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-23 UNCLASSIFIED The VERTICAL-DEFLECTION AMPLIFIER increases the amplitude of the input signal level required for the deflection of the CRT beam. The deflection amplifier must not have any other effect on the signal, such as changing the shape (called DISTORTION). Figure 6-21 shows the results of distortion occurring in a deflection amplifier.

6.3.3.3 Attenuator Control An amplifier can handle only a limited range of input amplitudes before it begins to distort the signal. Signal distortion is prevented in oscilloscopes by the incorporation of circuitry that permits adjustment of the input signal amplitude to a level that prevents distortion from occurring. This adjustment is called the ATTENUATOR control in some scopes (VOLTS/DIV and VAR in figure 6-19). This control extends the usefulness of the oscilloscope by enabling it to handle a wide range of signal amplitudes.

The attenuator usually consists of two controls. One is a multi-position (VOLTS/DIV) control, and the other is a variable (VAR) potentiometer. Each position of the control may be marked either as to the amount of voltage required to deflect the beam a unit distance, such as VOLTS/DIV, or as to the amount of attenuation (called the DEFLECTION FACTOR) given to the signal, such as 100, 10, or 1.

Figure 6-21 Deflection amplifier distortion

p. 249

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-24 UNCLASSIFIED Suppose the .5 VOLTS/DIV position were selected. In this position, the beam would deflect vertically 1 division for every 0.5 volts of applied signal. If a sine wave occupied 4 divisions peak-to-peak, its amplitude would be 2 volts peak-to-peak (4 × 0.5), as shown in figure 6-22.

The vertical attenuator control (VOLTS/DIV in figure 6-19) provides a means of adjusting the input signal level to the amplifiers by steps. These steps are sequenced from low to high deflection factors. The potentiometer control (VAR in figure 6-19) provides a means of fine, or variable, control between steps. This control may be mounted separately, or it may be mounted on the attenuator control. When the control is mounted separately, it is often marked as FINE GAIN or simply GAIN. When mounted on the attenuator control, it is usually marked VARIABLE or VAR.

The variable control adds attenuation to the step that is selected. Since accurately calibrating a potentiometer is difficult, the variable control is either left unmarked or the front panel is marked off in some convenient units, such as 1-10 and 1-100. The attenuator control, however, can be accurately calibrated. To do this, you turn off the variable control to remove it from the attenuator circuit. This position is usually marked CAL (calibrate) on the panel, or an associated light indicates if the VAR control is on or off. In figure 6-19, the light called UNCAL indicates the VAR control is in the uncalibrated position.

Figure 6-22 Sine wave attenuation

p. 250

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-25 UNCLASSIFIED 6.3.4 Components Used to Select the Vertical Operating Mode As we discussed earlier, channel 1 is being used to discuss basic operating procedures for the oscilloscope. Figure 6-23 shows how the vertical mode of operation is selected. The VERT MODE section contains push-button switches that enable you to select channel 1, channel 2, and several other vertical modes of operation. For the present discussion, note only that CH 1 is selected by these switches.

Figure 6-23 Vertical-deflection controls

p. 251

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-26 UNCLASSIFIED 6.3.5 Components Used to Determine Period Time of the Display The TIME/DIV (figure 6-24) controls on the scope determine the period time of the displayed waveform. As we discussed earlier, the sweep generator develops the sawtooth waveform that is applied to the horizontal-deflection plates of the CRT. This sawtooth voltage causes the beam to move across the screen. This trace (sometimes called SWEEP) sets the frequency of the TIME BASE of the oscilloscope. The frequency of the time base is variable, which enables the oscilloscope to accept a wide range of input frequencies. Again, two controls are used (figure 6-24). One is a multi-position switch (TIME/DIV) that changes the frequency of the sweep generator in steps. The second control is a potentiometer (VAR) that varies the frequency between steps. Each step on the TIME/DIV control is calibrated. The front panel has markings that group the numbers into microseconds and milliseconds.

Figure 6-24 Period time of the waveform (TIME/DIV)

p. 252

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-27 UNCLASSIFIED The potentiometer is labeled VAR, and the panel has an UNCAL indicator that lights when the VAR control is in the variable position. When you desire to accurately measure the time of one cycle of an input signal, turn the VAR control to the CAL position and turn the TIME/DIV switch to select an appropriate time base. Suppose you choose the 10-microsecond position to display two cycles of an input signal, as shown in figure 6-25. One cycle occupies 3 centimeters (small divisions) along the horizontal axis. Each cm has a value of 10 microseconds. Therefore, the time for one cycle equals 30 microseconds (3 × 10). Recall that the frequency for a signal may be found by using the following procedure:

f = 1 time (t)

= 1 30 X 10−6

= 33.33 kHz

In selecting a time base, you should select one that is lower in frequency than the input signal. If the input signal requires 5 milliseconds to complete one cycle and the sawtooth is set for 0.5 milliseconds per centimeter with a 10-centimeter-wide graticule, then approximately one cycle will be displayed. If the time base is set for 1 millisecond per centimeter, approximately two cycles will be displayed. If the time base is set at a frequency higher than the input frequency, only a portion of the input signal will be displayed.

Figure 6-25 Time measurement of a waveform (TIME/DIV)

p. 253

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-28 UNCLASSIFIED In the basic oscilloscope, the sweep generator runs continuously (FREE-RUNNING); in more elaborate oscilloscopes, it is normally turned off. In the oscilloscope we’re using as an example, the sweep generator can be triggered by the input signal or by a signal from some other source. (Triggering will be discussed later in this chapter.) This type of oscilloscope is called a triggered oscilloscope. The triggered oscilloscope permits more accurate time measurements to be made and provides a more stable presentation than the non-triggered-type oscilloscope.

On some oscilloscopes, you will find a 10 times (10X) magnification control. As previously mentioned, this allows the displayed sweep to be magnified by a factor of 10.

Q-15. When you select the time base to display a signal, should the time base be the same, higher, or lower than the input signal? 6.3.6 Components Used to Provide a Stable Display The triggering and level controls are used to synchronize the sweep generator with the input signal. This provides a stationary waveform display. If the input signal and horizontal sweep generator are unsynchronized, the pattern tends to jitter, making observations difficult.

The A TRIGGER controls at the lower right of the scope (figure 6-26) are used to control the stability of the oscilloscope CRT display. They are provided to permit you to select the source, polarity, and amplitude of the trigger signal. These controls, labeled A TRIGGER, LEVEL, SOURCE, and SLOPE, are described in the following paragraphs.

p. 254

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-29 UNCLASSIFIED

6.3.6.1 SOURCE Control The SOURCE control allows you to select the appropriate source of triggering. You can select input signals from channel 1 or 2, the line (60 hertz), or an external input.

Figure 6-26 Components that control stability

p. 255

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-30 UNCLASSIFIED 6.3.6.2 TRIGGER LEVEL/SLOPE Controls The LEVEL control allows you to select the amplitude point of the trigger signal at which the sweep is triggered. The SLOPE lets you select the negative or positive slope of the trigger signal at which the sweep is triggered. The TRIGGER LEVEL (mounted with the TRIGGER SLOPE on some scopes) determines the voltage level required to trigger the sweep. For example, in the TRIGGER modes, the trigger is obtained from the signal to be displayed. The setting of the LEVEL control determines the amplitude point of the input waveform that will be displayed at the start of the sweep.

Figure 6-27 shows some of the displays for a channel that can be obtained for different TRIGGER LEVEL and TRIGGER SLOPE settings. The level is zero and the slope is positive in view A; view B also shows a zero level but a negative slope selection. View C shows the effects of a positive trigger level setting and positive trigger slope setting; view D displays a negative trigger level setting with a positive trigger slope setting. Views E and F have negative slope settings. The difference is that view E has a positive trigger level setting, whereas F has a negative trigger level setting.

p. 256

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-31 UNCLASSIFIED

In most scopes, an automatic function of the trigger circuitry allows a free-running trace without a trigger signal. However, when a trigger signal is applied, the circuit reverts to the triggered mode of operation and the sweep is no longer free running. This action provides a trace when no signal is applied.

Synchronization is also used to cause a free-running condition without a trigger signal. Synchronization is not the same as triggering. TRIGGERING refers to a specific action or event that initiates an operation. Without this event, the operation would not occur. In the case of the triggered sweep, the sweep will not be started until a trigger is applied. Each succeeding sweep must have a trigger before a sweep commences. SYNCHRONIZATION, however, means that an operation or event is brought into step with a second operation. Figure 6-27 Effects of SLOPE and TRIGGER LEVEL controls

p. 257

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-32 UNCLASSIFIED A sweep circuit that uses synchronization instead of triggering will cause a previously free-running sweep to be locked in step with the synchronizing signal. The TRIGGER LEVEL control setting can be increased until synchronization occurs; but, until that time, an unstable pattern will appear on the CRT face.

6.3.6.3 COUPLING Section The COUPLING section allows you to select from four positions: AC, LF REJ, HF REJ, and DC. The AC position incorporates a coupling capacitor to block any dc component. The LF and HF REJ positions reject low- and high-frequency components, respectively. The DC position provides direct coupling to the trigger circuits. This is useful when you wish to view only the LF or HF component of a signal.

6.3.7 Components Used to Select Scope Triggering The TRIG MODE section in figure 6-28 allows for automatic triggering or normal triggering. In AUTO (automatic), the triggering will be free-running in the absence of a proper trigger input or will trigger on the input signal at frequencies above 20 hertz. In NORM (normal), the vertical channel input will trigger the sweep.

p. 258

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-33 UNCLASSIFIED

Figure 6-28 Components to select triggering

p. 259

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-34 UNCLASSIFIED 6.3.8 Components Used to Select Horizontal-Deflection Mode For the present, notice only that the HORIZ DISPLAY (horizontal display) in figure 6-29 can be controlled by the TIME/DIV switch. Other switches in this section will be explained later in this chapter.

Figure 6-29 Components to select mode of horizontal deflection

p. 260

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-35 UNCLASSIFIED 6.3.9 Components Used to Calibrate the Probe of the Scope In figure 6-30, you can see the components used to calibrate the test probe on the scope. A 1-volt, 2-kilohertz square wave signal is provided for you to adjust the probe for an accurate square wave and to check the vertical gain of the scope. You adjust the probe with a screwdriver, as shown in the figure.

6.3.10 Similarities Among Oscilloscopes The oscilloscope you use may differ in some respects from the one just covered. Controls and circuits may be identified by different names. Many of the circuits will be designed differently. However, all the functions will be fundamentally the same. Before using an oscilloscope, you should carefully study the operator’s manual that comes with it.

Figure 6-30 Components to calibrate probe

p. 261

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-36 UNCLASSIFIED 6.4 USING THE OSCILLOSCOPE An oscilloscope can be used for several different types of measurements, such as time, phase, frequency, and amplitude of observed waveforms. Earlier in this chapter, you learned that the oscilloscope is most often used to study the shapes of waveforms when the performance of equipment is being checked. The patterns on the scope are compared with the signals that should appear at test points (according to the technical manual for the equipment under test). You can then determine if the equipment is operating according to peak performance standards.

Q-16. Oscilloscopes are used to measure what quantities? 6.4.1 Turning on the Scope Before turning on the scope, make sure it is plugged into the proper power source. This may seem obvious, but many technicians have turned all knobs on the front panel out of adjustment before they noticed that the power cord was not plugged in. On some scopes, the POWER switch is part of the INTEN (intensity) control. Turn or pull the knob until you hear a click or a panel light comes on (figure 6-31). Let the scope warm up for a few minutes so that voltages in all of the circuits become stabilized.

Figure 6-31 Components to energize scope

p. 262

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-37 UNCLASSIFIED 6.4.2 Obtaining a Pattern on the Screen When adjusting a pattern onto the screen, adjust the INTEN (intensity) and FOCUS controls for a bright, sharp line. If other control settings are such that a dot instead of a line appears, turn down the intensity to prevent burning a hole in the screen coating. Because of the different speeds at which the beam travels across the screen, brightness and sharpness will vary at various frequency settings. For this reason, you may have to adjust the INTEN and FOCUS controls occasionally while taking readings.

6.4.3 Number of Cycles on the Screen Because distortion may exist at the beginning and end of a sweep, it is better to place two or three cycles of the waveform on the screen instead of just one, as shown in figure 6-32.

Figure 6-32 Proper signal presentation

p. 263

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-38 UNCLASSIFIED The center cycle of three cycles provides you with an undistorted waveform in its correct phase. The center of a two-cycle presentation will appear inverted, but will be undistorted. To place waveforms on the CRT in this manner, you must understand the relationship between horizontal and vertical frequencies. The relationship between the frequencies of the waveform on the vertical plates and the sawtooth on the horizontal plates determines the number of cycles on the screen, as shown in figure 6-33.

The horizontal sweep frequency of the scope should always be kept lower than, or equal to, the waveform frequency; it should never be higher. If the sweep frequency were higher, only a portion of the waveform would be presented on the screen.

If, for example, three cycles of the waveform were to be displayed on the screen, the sweep frequency would be set to one-third the frequency of the input signal. If the input frequency were 12,000 hertz, the sweep frequency would be set at 4,000 hertz for a three- cycle scope presentation. For two cycles, the sweep frequency would be set at 6,000 hertz. If a single cycle were desired, the setting would be the same as the input frequency, 12,000 hertz.

Figure 6-33 Vertical versus horizontal relationship

p. 264

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-39 UNCLASSIFIED 6.4.4 Dual-Trace Operation The information presented in the previous sections served as a general overview of basic single-trace oscilloscope operation using one channel and operating controls. Now, you will be introduced to DUALTRACE operation.

Dual-trace operation allows you to view two independent signal sources as a dual display on a single CRT. This operation allows an accurate means of making amplitude, phase, time displacement, or frequency comparisons and measurements between two signals.

A dual-trace oscilloscope should not be confused with a dual-beam oscilloscope. Dual- beam oscilloscopes produce two separate electron beams on a single scope, which can be individually or jointly controlled. Dual-trace refers to a single beam in a CRT that is shared by two channels.

Q-17. Scopes that produce two channels on a single CRT with a single beam are referred to as what types of scopes?

p. 265

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-40 UNCLASSIFIED 6.4.4.1 Components Used to Select Vertical-Deflection Operating Mode The VERT MODE controls (figure 6-34) allow you to select the operating mode of the scope for vertical deflection.

CH 1 AND CH 2 - These controls allow you to display signals applied to either channel 1 or channel 2, as discussed earlier.

TRIGGER VIEW - The TRIG VIEW allows you to display the signal that is actually used to trigger the display. (Triggering was discussed earlier.)

Figure 6-34 Components to select vertical operating mode

p. 266

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-41 UNCLASSIFIED ALT - The ALT (alternate) mode (figure 6-35) of obtaining a dual trace uses the techniques of GATING between sweeps. This control allows the signal applied to channel 1 to be displayed in its entirety; then, channel 2 is displayed in its entirety. This method of display is continued alternately between the two channels. At slow speeds, one trace begins to fade while the other channel is being gated. Consequently, the ALT mode is not used for slow sweep speeds. The CHOP mode, shown in figure 6-36 (explained next), will not produce a satisfactory dual sweep at high speeds. The ALT mode is deficient at low speeds. Therefore, both are used on dual-trace oscilloscopes to complement each other and give the scope a more dynamic range of operation.

Figure 6-35 ALT (alternate) mode Figure 6-36 CHOP mode

p. 267

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-42 UNCLASSIFIED The output dc voltage references on each of the amplifiers are independently adjustable. Therefore, the beam will be deflected by different amounts on each channel if the voltage reference is different at each amplifier output. The output voltage from each amplifier is applied to the deflection plates through the gate. The gate is actually an electronic switch. In this application, it is commonly referred to as a BEAM SWITCH.

Switching is controlled by a high-frequency multivibrator in the CHOP mode. That is, the gate selects the output of one channel and then the other at a high-frequency rate (1200 kilohertz in most oscilloscopes). Because the switching time is very short in a good- quality oscilloscope, the resultant display is two sets of horizontally dashed lines, as shown in figure 6-37, view A.

Figure 6-37 Displaying CHOP mode

p. 268

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-43 UNCLASSIFIED Dashed line CH 1 is the output of one channel, while line CH 2 is the output of the other. The trace moves from left to right because of the sawtooth waveform applied to the horizontal plates. A more detailed analysis shows that the beam moves from CH 1 to CH 2 while the gate is connected to the output from one channel. Then, when the gate samples the output of the CH 2 during time 3 to 4, the beam is at a different vertical LOCATION. (This is assuming that CH 2 is at a different voltage reference.) The beam continues in the sequence 5 to 6, 7 to 8, 9 to 10, and 11 to 12 through the rest of one horizontal sweep.

When the chopping frequency is much higher than the horizontal sweep frequency, the number of dashes will be very large. For example, if the chopping occurs at 100 kilohertz and the sweep frequency is 1 kilohertz, each horizontal line would then appear as a series of closely spaced dots, as shown in figure 6-37 view B. As the sweep frequency becomes lower compared to the chopping frequency, the display will show apparently continuous traces; therefore, the CHOP mode is used at low sweep rates.

When signals are applied to the channel amplifiers (view A of figure 6-38), the outputs are changed according to the triggering signal (view B). The resultant pattern (view C) on the screen provides a time-base presentation of the signals of each channel.

p. 269

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-44 UNCLASSIFIED

ADD - The ADD switch (shown earlier in figure 6-34) algebraically adds the two signals of channels 1 and 2 together for display.

Figure 6-38 Dual-channel display in CHOP mode

p. 270

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-45 UNCLASSIFIED 6.4.4.2 Other Dual-Trace Oscilloscope Controls Most dual-trace oscilloscopes have both an A and B time base for horizontal sweep control. Notice in the upper right corner on our example scope (figure 6-34) the COUPLING, SOURCE LEVEL, and SLOPE controls. These serve the same function as did those same controls in the A time-base section of the scope. The B time base is selected using the same A and B TIME/DIV control (pull out outer knob).

The use of the B time base is controlled by the HORIZ DISPLAY section discussed earlier in the A time-base section. However, inexperienced technicians generally do not use A and B time bases together in the MIXED, A INTEN (intensified), and B D'LYD (delayed) settings. These controls are fully explained in the applicable technical manuals; therefore, we will not discuss the controls in this chapter. Figure 6-39 is a block diagram of a basic dual-trace oscilloscope without the power supplies.

Figure 6-39 Basic dual-trace oscilloscope block diagram

p. 271

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-46 UNCLASSIFIED 6.4.5 Accessories The basic dual-trace oscilloscope has one gun assembly and two vertical channels. However, there are many variations. The horizontal sweep channels vary somewhat from equipment to equipment. Some have one time-base circuit and others have two. These two are interdependent in some oscilloscopes and in others they are independently controlled. Also, most modern general-purpose oscilloscopes are constructed of modules. That is, most of the vertical circuitry is contained in a removable plug-in unit, and most of the horizontal circuitry is contained in another plug-in unit.

The main frame of the oscilloscope is often adapted for many other special applications by the design of a variety of plug-in assemblies. This modular feature provides much greater versatility than in a single-trace oscilloscope. For instance, to analyze the characteristics of a transistor, you can replace the dual-trace, plug-in module with a semiconductor curve-tracer plug-in module.

Other plug-in modules available with some oscilloscopes are high-gain, wide-bandwidth amplifiers; differential amplifiers; spectrum analyzers; physiological monitors; and other specialized units. Therefore, the dual-trace capability is a function of the type of plug-in unit that is used with some oscilloscopes.

To get maximum usefulness from an oscilloscope, you must have a means of connecting the desired signal to the oscilloscope input. Aside from cable connections between any equipment output and the oscilloscope input, a variety of probes are available to assist in monitoring signals at almost any point in a circuit. The more common types include 1- TO-1 PROBES, ATTENUATION PROBES, and CURRENT PROBES. Each of these probes may be supplied with several different tips to allow measurement of signals on any type of test point. Figure 6-40 shows some of the more common probe tips.

In choosing the probe to use for a particular measurement, you must consider such factors as circuit loading, signal amplitude, and scope sensitivity.

The 1-to-1 probe offers little or no attenuation of the signal under test and is, therefore, useful for measuring low-level signals. However, circuit loading with the 1-to-1 probe may be a problem. The impedance at the probe tip is the same as the input impedance of the oscilloscope.

p. 272

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-47 UNCLASSIFIED

An attenuator probe has an internal high-value resistor in series with the probe tip. This gives the probe a higher input impedance than that of the oscilloscope. Because of the higher input impedance, the probe can measure high-amplitude signals that would overdrive the vertical amplifier if connected directly to the oscilloscope. Figure 6-41 shows a schematic representation of a basic attenuation probe. The 9-megohm resistor in the probe and the 1-megohm input resistor of the oscilloscope form a 10-to-1 voltage divider.

Figure 6-40 Common probe tips Figure 6-41 Basic attenuation probe

p. 273

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-48 UNCLASSIFIED Since the probe resistor is in series, the oscilloscope input resistance is 10 megohms when the probe is used. Thus, using the attenuator probe with the oscilloscope causes less circuit loading than using a 1-to-1 probe.

Before using an attenuator probe for measurement of high-frequency signals or for fast- rising waveforms, you must adjust the probe compensating capacitor (C1) according to instructions in the applicable technical manual. Some probes will have an IMPEDANCE EQUALIZER in the end of the cable that attaches to the oscilloscope. The impedance equalizer, when adjusted according to manufacturer’s instructions, assures proper impedance matching between the probe and oscilloscope. An improperly adjusted impedance equalizer will result in erroneous measurements, especially when you are measuring high frequencies or fast-rising signals.

More information on oscilloscope hook-ups can be found in Electronics Information Maintenance Books (EIMB), Test Methods and Practices.

Special current probes have been designed to use the electromagnetic fields produced by a current as it travels through a conductor. This type of probe is clamped around a conductor without disconnecting it from the circuit. The current probe is electrically insulated from the conductor, but the magnetic fields about the conductor induce a potential in the current probe that is proportional to the current through the conductor. Thus, the vertical deflection of the oscilloscope display will be directly proportional to the current through the conductor.

p. 274

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-49 UNCLASSIFIED 6.5 SPECTRUM ANALYZER The spectrum analyzer is used to examine the frequency spectrum of radar transmissions, local oscillators, test sets, and any other equipment operating within its testable frequency range. With experience, you will be able to determine definite areas of malfunctioning components within equipment. Successful spectrum analysis depends on the proper operation of a spectrum analyzer and your ability to correctly interpret the displayed frequencies. Although there are many types of spectrum analyzers, we will use the Tektronix, Model 492 for our discussion.

The spectrum analyzer accepts an electrical input signal and displays the frequency and amplitude of the signal on a CRT. On the vertical, or Y, axis, the amplitude is plotted. The frequency would then be found on the horizontal, or X, axis. The overall pattern of this display (figure 6-42) indicates the proportion of power present at the various frequencies within the spectrum (fundamental frequency with sideband frequencies).

Figure 6-42 Spectrum analyzer pattern

p. 275

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-50 UNCLASSIFIED 6.5.1 Basic Functional Description The model 492 analyzer can be divided into six basic sections, as follows:

• Converter section; • Intermediate frequency (IF) section; • Display section; • Frequency control section; • Digital control section; and • Power and cooling section.

6.5.1.1 Converter Section The converter section actually consists of three frequency converters, made up of a mixer, local oscillator (LO), and required filters. Only one frequency can be converted at a time and pass through the filters to reach the next converter. The analysis frequency can, however, be changed by altering the frequency of the LO and adjusting the FREQUENCY control knob.

FIRST CONVERTER - The first (front end) converter changes the input signal to a usable IF signal that will either be 829 MHz or 2072 MHz. The IF signal to be produced is dependent on which measurement band selection is currently being used. The 829 MHz IF signal will be selected for bands 2 through 4, while the 2072 MHz IF signal is selected for bands 1 and 5 through 11.

Q-18. The first converter is also known by what other name? SECOND CONVERTER - The second converter actually contains two converters. Only one of these two converters in this section is ever operational, and selected as a result of the measurement band currently being used. The selected converter will convert the frequency received from the first converter to a usable (110 MHz) IF signal, which is then provided to the third converter.

THIRD CONVERTER - This converter takes the 110 MHz IF signal, amplifies it, and then converts it to the final IF of 10 MHz. This signal, in turn, is then passed on to the IF section.

p. 276

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-51 UNCLASSIFIED 6.5.1.2 IF Section The IF section receives the final IF signal and uses it to establish the system resolution by using selective filtering. System resolution is selected under microcomputer control among five bandwidths (1 MHz, 100 kHz, 10 kHz, 1 kHz, and 100 Hz). The gain for all bands are then leveled and logarithmically amplified. This is done so that each division of signal change on the CRT display remains equal in change to every other division on the CRT. For example, in the 10-dB-per-division mode, each division of change is equal to a 10 dB difference, regardless of whether the signal appears at the top or bottom of the CRT. The signal needed to produce the video output to the display section is then detected and provided.

6.5.1.3 Display Section The display section provides a representative display of the input signal on the CRT. It accomplishes this by performing the following functions:

• Receives the video signals from the IF section and processes these signals to adjust the vertical drive of the CRT;

• Receives the sweep voltages and processes these signals to produce the horizontal CRT drive plate voltage;

• Receives character data information and generates CRT plate drive signals to display alpha and numeric characters on the CRT;

• Receives control levels from the front panel beam controls and generates unblanking signals to control display presence, brightness, and focus.

The vertical deflection of the beam is increased as the output of the amplitude detector increases. The horizontal position is controlled by the frequency control section and is the frequency analyzed at that instant. The beam sweeps from left to right, low to high frequencies during its analysis. During this analysis, any time a signal is discovered, a vertical deflection will show the strength of the signal at the horizontal position that is the frequency. This results in a display of amplitude as a function of frequency.

p. 277

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-52 UNCLASSIFIED 6.5.1.4 Frequency Control Section The frequency control section accomplishes the tuning of the first and second LOs within the converter section. The frequency immediately being analyzed is controlled by the current frequencies of the LOs. To analyze another frequency, you must change an LO frequency to allow the new frequency to be converted to a 10 MHz signal by the converter section. Periodically, the unit sweeps and analyzes a frequency range centered on the frequency set by the FREQUENCY knob. Adjusting the FREQUENCY knob will cause the LOs to be tuned to the new frequency. Only the LOs of the first two converters can be changed to vary the frequency being analyzed.

6.5.1.5 Digital Control Section All the internal functions are controlled from the front panel through the use of a built-in microcomputer. The microcomputer uses an internal bus to receive or produce all communication or control to any section of the analyzer.

6.5.1.6 Power and Cooling Section The main power supply provides almost all the regulated voltages required to operate the unit. The display section provides the high voltage necessary for CRT operation.

The cooling system allows fresh cool air to be routed to all sections of the unit in proportion to the heat that is generated by each section.

6.5.2 Spectrum Analyzer Front Panel Controls, Indicators, and Connectors This section will describe the function of the front panel controls, indicators, and connectors. For a complete description of each function, refer to table 6-1 while reviewing the front panel in figure 6-43. The numbers located in column 1 of table 6-1 equate to the same numbers found on the front panel of figure 6-43. Because most operational functions of this spectrum analyzer are microprocessor-controlled, they are switch-selected rather than adjusted.

p. 278

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-53 UNCLASSIFIED

Figure 6-43 Spectrum analyzer front panel controls, indicators, and connectors

p. 279

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-54 UNCLASSIFIED Table 6-1 Description of Front Panel Controls, Indicators, and Connectors ITEM FUNCTION DESCRIPTION 1 INTENSITY This knob controls the brightness of the CRT trace and the CRT readout display. The focus is electronically adjusted. 2 READOUT This push button switches the readout display on and off. All spectrum analyzer parameters are displayed except TIME/DIV. The brightness for this display is proportional to the trace brightness and can be readjusted on internal controls only by a qualified technician. 3 GRATILLUM This push button switches the graticule light on and off. 4 BASELINE CLIP This push button, when activated, clips (subdues) the intensity at the baseline. 5 TRIGGERING This area allows one of four triggering modes to be selected by push buttons that illuminate when active. When any of these four are selected, the others are canceled. 5a FREE RUN When activated, the sweep is free-running without regard to trigger signals. 5b INT When activated, the sweep is triggered by any signal at the left edge of the display with an amplitude of 1.0 divisions of the graticule or more. 5c LINE When activated, a sample of the ac power line voltage is used to trigger the sweep. 5d EXT When selected, the sweep is triggered by an external signal (applied through the back panel IN HORZ/TRIG connector) between a minimum and maximum of 0.5 and 50 volt peak. 6 SINGLE SWEEP This push button, plus a ready indicator (No. 7), provides the single sweep operation. When this operation is selected, one sweep is initiated after the sweep circuit has been triggered. Pushing this button does not cancel the other trigger modes. When single sweep is first selected, the present sweep is aborted, but the sweep circuit is not yet armed. An additional push is required to initially arm the sweep. The button must be pushed again to rearm the sweep circuit each time the sweep has run. To cancel single sweep, you must select one of the four trigger mode selections. 7 READY When single sweep is selected, this indicator lights while the sweep circuit is armed and ready for a trigger signal. The indicator stays lit until the sweep is complete. 8 MANUAL SCAN When the TIME/DIV (No. 9c) selector is in the MNL position, this control will manually vary the CRT beam across the full horizontal axis of the display. 9 TIME/DIV Is used to select sweep rates from 5μsec/div to 20μsec/div. This switch also selects AUTP, EXT, and MNL modes. 9a AUTO In this position, the sweep rate is selected by the microcomputer to maintain a calibrated display for any FREQ SPAN/DIV, RESOLUTION, and VIDEO FILTER combination. 9b EXT When selected, this control allows an external input source to be used with the sweep rates. 9c MNL When selected, this control is used in conjunction with No. 8 (see MANUAL SCAN, No. 8). 10 FREQUENCY This control is manually turned to allow you to tune to the center frequency. 11 FREQUENCY RANGE (band) These two push buttons are used to shift the center frequency up or down. Frequency range on the band is displayed on the CRT readout.

p. 280

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-55 UNCLASSIFIED ITEM FUNCTION DESCRIPTION 12 F This control is used for measuring the frequency difference between signals. When selected, the frequency readout goes to zero. It will then read out the deviation from this reference to the next frequency desired as the FREQUENCY knob is adjusted. 13 CAL When this is activated, the frequency readout can be calibrated to center the center frequency by adjusting the FREQUENCY control for the correct reading. When accomplished, you should deactivate the CAL mode. 14 DEGAUSS When this button is pressed, current through the local oscillator system is reduced to zero in order to minimize magnetism build-up around the LOs. This is done to enhance the center frequency display and amplitude accuracy. You should do this after every significant frequency change and before calibrating the center frequency. 15 IDENTIFY 500 kHz ONLY The signal identify feature can become functional only when the FREQ SPAN/DIV is set to 500 kHz. When activated (button lit), true signals will change in amplitude on every sweep. Images and spurious response signals will shift horizontally or go completely off the CRT display. To ensure that the signal is changing amplitude every sweep, you should decrease the sweep rate so that each sweep can be analyzed. 16 PHASE LOCK When this control is activated (button lit), it will reduce residual FM when narrow spans are selected. In narrow spans, the phase lock can be turned off or back on by pressing the button. Switching the PHASE LOCK off may cause the signal to shift position. In narrow spans, the signal could shift off the display; however, it will usually return to its phase locked position after a few moments. The microcomputer automatically selects PHASE LOCK for a span/division of 50 kHz or below in bands 1 through 3, 100 kHz or below for band 3, and 200 kHz for bands 5 and above. 17 AUTO RESOLUTION This push button, when activated, will automatically select the bandwidth for FREQ SPAN/DIV, TIME/DIV, and VIDEO FILTER. The internal microcomputer selects the bandwidth to maintain a calibrated display. This can be checked by changing the FREQ SPAN/DIV and observing the bandwidth change on the display. 18 FREQ SPAN/DIV This is a continuous detent control that selects the frequency span/div. The span/div currently selected is displayed on the CRT. The range of the span/div selection is dependent on the frequency band selected:

BAND NARROW SPAN WIDE SPAN 1-3 (0-7.1GHz) 10kHz/Div 200MHz/Div 4-5 (5.4-21GHz) 50 kHz/Div 500 MHz/Div 6 (18-26GHz) 50 kHz/Div 1 GHz/Div 7-8 (26-60GHz) 100 kHz/Div 2 GHz/Div 9 (60-90GHz) 200 kHz/Div 2 GHz/Div 10 (90-140GHz) 500 kHz/Div 5 GHz/Div 11 (140-220GHz) 500 kHz/Div 10 GHz/Div Two additional bands are provided: full band (max span) and 0 Hz span. When max span is selected, the span displayed is the full band. When zero span is selected, time/div is read out instead of span/div.

p. 281

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-56 UNCLASSIFIED ITEM FUNCTION DESCRIPTION 19 RESOLUTION BANDWIDTH This is also a continuous detent control that selects the resolution bandwidth. The bandwidth is shown on the CRT display. The range of adjustment is from 1 kHz to 1 MHz in decade steps. When you change the resolution bandwidth with this control, it will deactivate the AUTO RESOLUTION. 20 VERTICAL DISPLAY These four push buttons select the display mode. The scale factor can be seen on the CRT display. 20a 10dB/DIV When this is activated, the dynamic range of the display is calibrated to 80 dB, with each major graticule representing 10 dB. 20b 2dB/DIV When activated, this will increase the resolution so that each major graticule division represents 2 dB. 20c LIN When activated, this selects a linear display between zero volts (bottom graticule line) and the reference level (top graticule line) scaled in volts/division (see REFERENCE LEVEL, No. 23a). 20d PULSE STRETCHER When selected, this increases the fall time of the pulse signals so that very narrow pulses in a line spectrum display can be observed. 21 VIDEO FILTER One of two (NARROW OR WIDE) filters can be activated to reduce video bandwidth and high-frequency components for display noise averaging. The narrow filter is approximately 1/300th of the selected resolution bandwidth with the wide filter being 1/30th the bandwidth. Activating either one will cancel the other. To disable, completely switch filter off. 22 DIGITAL STORAGE Five push buttons and ON control operate the digital storage functions. With none of the push buttons activated, the display will not be stored. 22a VIEW A, VIEW B When either or both of these push buttons are selected, the push button illuminates, and the contents of memory A and/or memory B are displayed. With Save A mode off, data in a memory is interlaced with data from B memory. 22b B-SAVE A When activated, the differential (arithmetic difference) of data in B memory and the saved data in memory A are displayed. SAVE A mode is activated and SAVE A button will be lit. 22c MAX HOLD When activated, the digital storage memory retains the maximum signal amplitude at each memory location. This permits visual monitoring of signal frequency and amplitude at each memory location over an indefinite period of time. This feature is used to measure drift, stability, and record peak amplitude. 22d PEAK/AVERAGE This control selects the amplitude at which the vertical display is either peak detected or averaged. Video signals above the level set by the control (shown by a horizontal line or cursor) are peak detected and stored while video signals below the cursor are digitally averaged and stored. 23 MIN RF ATTEN This control is used to set the minimum amount of RF attenuation. Changing RF LEVEL will not decrease RF attenuation below that set by the MIN RF ATTEN selector. 23a REFERENCE LEVEL This is a continuous control that requests the microcomputer to change the reference level one step for each detent. In the 10 dB/DIV vertical-display mode, the steps are 1 dB or 0.25 dB if the FINE mode (No. 26) is selected.

p. 282

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-57 UNCLASSIFIED ITEM FUNCTION DESCRIPTION 23b MIN RF ATTEN DB This selects the lowest value of attenuation allowed: Actual RF attenuation is set by the microcomputer according to the logarithm selected by the MIN NOISE/MIN DISTORTION (No. 27) button. If RF attenuation is increased by changing MIN RF ATTEN, the microcomputer automatically changes IF gain to maintain the current reference level. 24 UNCAL This indicator lights when the display amplitude is no longer calibrated (selecting a sweep rate that is not compatible with the frequency span/div and resolution bandwidth). 25 LOG and AMPL CAL These adjustments calibrate the dynamic range of the display. The LOG calibrates any logarithm gain dB/Div, and the AMPL calibrates the reference level of the top graticule line at the top of the display. 26 FINE When activated, the REFERENCE LEVEL (No. 23a) switches in 1 dB increments for 10 dB/Div display mode, 0.25 dB for 2 dB/Div, and volts 1 dB for LIN display mode. 27 MIN NOISE/MIN DISTORTION This selects one of two logarithms used to control attenuator and IF gain. MIN NOISE (button illuminated) reduces the noise level by reducing attenuation and IF gain 10 dB. MIN DISTORTION (button not illuminated) reduces distortion to its minimum. To observe any changes, the RF attenuation displayed on the CRT readout must be 10 dB higher than that set by the MIN RF ATTEN selector. 28 POWER This is a pull switch that turns power on when extended. 29 RF INPUT This is a 50 ohm coaxial input jack used to input signals of 21GHz or below. The maximum nondestructive input signal level that can be applied to this input is +13 dBm or 30 mW. Signals above 10 dB may cause signal compression. 30 POSITION These controls are used to position the display on the horizontal and vertical axes. 31 CAL OUT This is an output jack that has a calibrated 20 dBm 100 MHz signal, with frequency markers spaced 100 MHz apart. The calibrated 100 MHz marker is used as a reference for calibrating the reference level and log scale. The combination of 100 MHz markers is used to check span and frequency readout accuracy. 32 OUTPUT 1ST AND 2ND LO These jacks provide access to the output of the respective LOs. The jacks must have 50 ohm terminators installed when not connected to an external device. 33 EXTERNAL MIXER When the EXTERNAL MIXER button is activated, bias is provided out the EXTERNAL MIXER port for external waveguide mixers. The IF output from the EXTERNAL MIXER is then applied through the EXTERNAL MIXER port to the second converter for use. 34 PEAKING This control varies the mixer bias for external mixers in the EXTERNAL MIXER mode. This control should be adjusted for maximum signal amplitude.

p. 283

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-58 UNCLASSIFIED 6.5.3 Normal Indications Upon Power on With power applied (power knob pulled out), the spectrum analyzer will automatically (upon microcomputer control) go into the following conditions. If you do not find these indications, there is a probably a problem with the unit.

• Vertical display: 10 dB/div;

• Frequency: 0.00 MHz;

• REF level: +30 dB;

• RF attenuation: 60 dB;

• Frequency range: 0.0 to 1.8 GHz;

• Auto resolution: 1 MHz;

• Resolution bandwidth: 1 MHz;

• Freq Span/Div: Max;

• Triggering: Free run;

• Readout: On;

• Digital storage: View A/View B On;

• All other indicators off or inactive.

p. 284

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-59 UNCLASSIFIED 6.6 SUMMARY Now that we have completed this chapter, we will briefly review the more important points covered.

A CATHODE-RAY TUBE (CRT) is used in an oscilloscope to display the waveforms.

The CRT used in oscilloscopes consists of an ELECTRON GUN, a DEFLECTION SYSTEM, and a FLUORESCENT SCREEN.

The ELECTRON BEAM in an oscilloscope is allowed to be controlled in any direction by means of HORIZONTAL- and VERTICAL-DEFLECTION PLATES.

VERTICAL-DEFLECTION PLATES are used to show AMPLITUDE of a signal.

HORIZONTAL-DEFLECTION PLATES are used to show TIME and/or FREQUENCY relationship.

A GRATICULE is a calibrated scale of AMPLITUDE VERSUS TIME that is placed on the face of the CRT.

A DUAL-TRACE OSCILLOSCOPE is designed to accept two vertical inputs at the same time. It uses a single beam of electrons shared by two channels.

The SPECTRUM ANALYZER accepts an electrical input signal and displays the signal’s frequency and amplitude on a CRT display.

p. 285

NEETS Module 16, NAVEDTRA 14188A UNCLASSIFIED

6-60 UNCLASSIFIED ANSWERS TO QUESTIONS Q1. THROUGH Q18.

A-1. Control grid.

A-2. The first anode.

A-3. Because they bend electron streams in much the same manner that optical lenses bend light rays.

A-4. It accelerates the electrons emerging from the first anode.

A-5. A greater deflection angle.

A-6. A greater deflection angle.

A-7. Higher potential.

A-8. Slower beam.

A-9. Amplitude and time.

A-10. Amplitude.

A-11. Time and/or frequency relationships.

A-12. To permit wide-angle deflection of the beam.

A-13. Deflection factor.

A-14. A CRT, a group of control circuits, power supply, sweep circuitry, and deflection circuitry.

A-15. Lower.

A-16. Amplitude, phase, time, and frequency.

A-17. Dual-trace oscilloscopes.

A-18. Front end.